Arthroscopy. 2010 Nov;26(11):1474-7. doi: 10.1016/j.arthro.2010.02.027. Epub 2010 Sep 20.
The purpose of this study was to determine whether an all-inside peripheral triangular fibrocartilage complex (TFCC) repair using the FasT-Fix device (Smith & Nephew Endoscopy, Andover, MA) is safe by measuring the proximity of the anchors to ulnar-sided anatomic structures.
Eleven fresh-frozen cadaveric wrists were thawed and placed in traction. Under direct arthroscopic visualization, an all-inside arthroscopic peripheral TFCC repair was completed by placing a single FasT-Fix device in a vertical mattress fashion. The wrists were then dissected to visualize the 2 anchors. The distance between these anchors and the flexor carpi ulnaris (FCU), extensor carpi ulnaris (ECU), and dorsal branch of the ulnar sensory nerve (DBUN) were measured with digital calipers and recorded.
The peripheral anchor averaged 4.2 mm (range, 0 to 14 mm) from the ECU tendon, 3.8 mm (range, 0 to 9 mm) from the DBUN, and 8.3 mm (range, 1 to 15 mm) from the FCU tendon. The central anchor averaged 9.6 mm (range, 2 to 15 mm) from the ECU tendon, 6.8 mm (range, 1 to 13 mm) from the DBUN, and 7.6 mm (range, 1 to 13 mm) from the FCU tendon.
This study exposes some safety concerns with the all-inside peripheral TFCC repair using the FasT-Fix device, which was found to reside in close proximity to the ECU, FCU, and DBUN. In multiple wrists the anchors were noted to underlie the anatomic structure that we measured, making it possible to pierce these structures with the needle before deployment of the anchor.
Though technically feasible, all-inside arthroscopic repair of the peripheral TFCC risks injury to the ulnar-sided anatomy.
本研究旨在通过测量锚钉与尺侧解剖结构的接近程度,确定使用 FasT-Fix 装置(Smith & Nephew Endoscopy,马萨诸塞州安多弗)进行全内缘周围三角纤维软骨复合体(TFCC)修复是否安全。
11 个新鲜冷冻的尸体手腕解冻后置于牵引下。在直接关节镜可视化下,通过以垂直褥式方式放置单个 FasT-Fix 装置完成全内缘关节镜下周围 TFCC 修复。然后解剖这些手腕以可视化 2 个锚钉。使用数字卡尺测量这些锚钉与尺侧腕屈肌(FCU)、尺侧腕伸肌(ECU)和尺神经背侧支(DBUN)之间的距离,并记录下来。
外周锚钉平均距离 ECU 肌腱 4.2 毫米(范围 0 至 14 毫米),距离 DBUN 3.8 毫米(范围 0 至 9 毫米),距离 FCU 肌腱 8.3 毫米(范围 1 至 15 毫米)。中央锚钉平均距离 ECU 肌腱 9.6 毫米(范围 2 至 15 毫米),距离 DBUN 6.8 毫米(范围 1 至 13 毫米),距离 FCU 肌腱 7.6 毫米(范围 1 至 13 毫米)。
这项研究揭示了使用 FasT-Fix 装置进行全内缘周围 TFCC 修复存在一些安全隐患,发现该装置与 ECU、FCU 和 DBUN 非常接近。在多个手腕中,我们发现锚钉位于我们测量的解剖结构下方,因此在部署锚钉之前,有可能刺穿这些结构。
尽管技术上可行,但全内缘关节镜修复周围 TFCC 可能会损伤尺侧解剖结构。