使用呼吸管时摄氧量动力学是否会改变?

Are oxygen uptake kinetics modified when using a respiratory snorkel?

机构信息

Department of Sport Sciences, Faculty of Human Kinetics, Technical University of Lisbon, Portugal.

出版信息

Int J Sports Physiol Perform. 2010 Sep;5(3):292-300. doi: 10.1123/ijspp.5.3.292.

Abstract

PURPOSE

The aim of this study was to compare VO2 kinetics during constant power cycle exercise measured using a conventional facemask (CM) or a respiratory snorkel (RS) designed for breath-by-breath analysis in swimming.

METHODS

VO2 kinetics parameters-obtained using CM or RS, in randomized counterbalanced order-were compared in 10 trained triathletes performing two submaximal heavy-intensity cycling square-wave transitions. These VO2 kinetics parameters (ie, time delay: td1, td2; time constant: τ1, τ2; amplitude: A1, A2, for the primary phase and slow component, respectively) were modeled using a double exponential function. In the case of the RS data, this model incorporated an individually determined snorkel delay (ISD).

RESULTS

Only td1 (8.9 ± 3.0 vs 13.8 ± 1.8 s, P < .01) differed between CM and RS, whereas all other parameters were not different (τ1 = 24.7 ± 7.6 vs 21.1 ± 6.3 s; A1 = 39.4 ± 5.3 vs 36.8 ± 5.1 mL x min(-1) x kg(-1); td2 = 107.5 ± 87.4 vs 183.5 ± 75.9 s; A2' (relevant slow component amplitude) = 2.6 ± 2.4 vs 3.1 ± 2.6 mL x min(-1) x kg(-1) for CM and RS, respectively).

CONCLUSIONS

Although there can be a small mixture of breaths allowed by the volume of the snorkel in the transition to exercise, this does not appear to significantly influence the results. Therefore, given the use of an ISD, the RS is a valid instrument for the determination of VO2 kinetics within submaximal exercise.

摘要

目的

本研究旨在比较使用传统面罩(CM)或专为游泳呼吸分析而设计的呼吸管(RS)测量的恒功率循环运动中的 VO2 动力学。

方法

以随机交叉平衡的顺序,使用 CM 或 RS 获得的 VO2 动力学参数(即,时滞:td1、td2;时间常数:τ1、τ2;幅度:A1、A2,分别用于主相和慢分量)在 10 名接受过训练的三项全能运动员中进行比较,他们进行了两次次最大强度的自行车方波过渡。对于 RS 数据,该模型包含个体确定的呼吸管延迟(ISD)。

结果

仅 CM 和 RS 之间的 td1(8.9 ± 3.0 对 13.8 ± 1.8 s,P <.01)不同,而其他参数没有差异(τ1 = 24.7 ± 7.6 对 21.1 ± 6.3 s;A1 = 39.4 ± 5.3 对 36.8 ± 5.1 mL x min(-1) x kg(-1);td2 = 107.5 ± 87.4 对 183.5 ± 75.9 s;A2'(相关的慢分量幅度)= 2.6 ± 2.4 对 3.1 ± 2.6 mL x min(-1) x kg(-1),分别为 CM 和 RS)。

结论

尽管呼吸管在过渡到运动时允许混合少量呼吸,但这似乎并没有显著影响结果。因此,在使用 ISD 的情况下,RS 是确定亚最大运动中 VO2 动力学的有效仪器。