猪体内电传导设备的心脏安全性及其对起搏器功能的影响。

Cardiac safety of conducted electrical devices in pigs and their effect on pacemaker function.

机构信息

Department of Internal Medicine, University of Missouri Columbia, Columbia, MO 65212, USA.

出版信息

Am J Emerg Med. 2011 Nov;29(9):1089-96. doi: 10.1016/j.ajem.2010.07.007. Epub 2010 Oct 12.

Abstract

OBJECTIVE

The aims of this study are to evaluate the cardiac safety of the Stinger S-200 Conducted Energy Weapon Device (CED) (Stinger Systems, Tampa, Fla) on a human-sized pig model and to test the effect of various commercially available CEDs, specifically the Stinger S-200, TASER M26 (Taser International, Scottsdale, Ariz), and TASER X26 on pacemaker function.

METHODS

Two groups of pigs, divided based on weight as group 1 (n = 3, 67.3 ± 4.7 kg) and group 2 (n = 3, 89.3 ± 1.2 kg), were used. In protocol 1, the Stinger S-200 was applied in multiple different orientations to simulate possible field scenarios across the heart. In protocol 2, a single-chamber bipolar lead connected to a pacemaker was placed in the right ventricle of the pig, and different CEDs were applied to test the pacemaker function during CED application.

RESULTS

In protocol 1, the S-200 was applied a total of 216 times in the 6 pigs, and neither episodes of ventricular fibrillation nor episodes of sustained ventricular tachycardia were noted. In protocol 2, the CED discharges (1) were recognized by the pulse generator and sensed as either high-rate atrial or ventricular activity, (2) did not affect the native rhythm, (3) did not conduct down the lead systems to cause any extra systoles, and (4) had no effect on paced rhythm.

CONCLUSIONS

In this model, the application of the S-200 in various orientations across the heart did not result in any sustained abnormal cardiac rhythms. None of the tested CEDs adversely affected the functioning of the tested pacemaker. Stinger Systems has now replaced the S-200 with the S-200T with a different output.

摘要

目的

本研究旨在评估 Stinger S-200 传导能武器设备(CED)(Stinger Systems,佛罗里达州坦帕)在人体大小猪模型中的心脏安全性,并测试各种市售 CED 的效果,特别是 Stinger S-200、TASER M26(Taser International,亚利桑那州斯科茨代尔)和 TASER X26 对起搏器功能的影响。

方法

使用两组猪,根据体重分为第 1 组(n = 3,67.3 ± 4.7 kg)和第 2 组(n = 3,89.3 ± 1.2 kg)。在方案 1 中,将 Stinger S-200 以多种不同的方向应用于模拟心脏的各种可能的现场场景。在方案 2 中,将一个连接到起搏器的单腔双极导联放置在猪的右心室中,并应用不同的 CED 来测试 CED 应用期间的起搏器功能。

结果

在方案 1 中,在 6 只猪中总共应用了 216 次 S-200,未出现心室颤动或持续室性心动过速发作。在方案 2 中,CED 放电(1)被脉冲发生器识别并感知为高心率心房或心室活动,(2)不影响固有节律,(3)不通过导联系统传导导致任何额外的收缩,(4)对起搏节律没有影响。

结论

在该模型中,将 S-200 以各种方向应用于心脏并未导致任何持续的异常心脏节律。测试的 CED 均未对测试的起搏器功能产生不利影响。Stinger Systems 现已用不同输出的 S-200T 取代了 S-200。