Performance of an automatic dose control system for CT: specifications and basic phantom tests.

作者信息

Nagel H D, Stumpp P, Kahn T, Gosch D

机构信息

Wissenschaft & Technik für die Radiologie, Buchholz, Germany.

出版信息

Rofo. 2011 Jan;183(1):60-7. doi: 10.1055/s-0029-1245721. Epub 2010 Oct 11.

Abstract

PURPOSE

To assess the performance and to provide more detailed insight into the characteristics and limitations of devices for automatic dose control (ADC) in CT.

MATERIALS AND METHODS

A comprehensive study on DoseRight 2.0, the ADC system provided by Philips for its Brilliance CT scanners, was conducted. Phantom tests were carried out on a 64-slice scanner (Brilliance 64) using assorted quality control (QC) phantoms that allowed verification of the basic specifications. If feasible, the findings were verified by model calculations based on known specifications.

RESULTS

For all tests, the dose reductions and modulation characteristics fully met the values expected from the specifications. Adverse effects due to increased image noise were only moderate as a result of the "adequate noise system" design that employs comparatively gentle modulation, and the additional use of adaptive filtration.

CONCLUSION

Simple tests with QC phantoms allow evaluation of the most relevant characteristics of devices for ADC in CT.

摘要

目的

评估CT中自动剂量控制(ADC)设备的性能,并更深入地了解其特性和局限性。

材料与方法

对飞利浦为其Brilliance CT扫描仪提供的ADC系统DoseRight 2.0进行了全面研究。在一台64层扫描仪(Brilliance 64)上使用各种质量控制(QC)模体进行模体测试,这些模体可用于验证基本规格。如果可行,通过基于已知规格的模型计算来验证结果。

结果

对于所有测试,剂量降低和调制特性完全符合规格预期值。由于采用相对温和调制的“适当噪声系统”设计以及额外使用自适应过滤,图像噪声增加所导致的不良影响仅为中等程度。

结论

使用QC模体进行简单测试可评估CT中ADC设备的最相关特性。