Performance and reliability of a variable rate, force/displacement application system.

作者信息

Vaillant Michèle, Pickar Joel G, Kawchuk Gregory N

机构信息

Graduate Student, University of Alberta, Edmonton, Alberta, Canada.

出版信息

J Manipulative Physiol Ther. 2010 Oct;33(8):585-93. doi: 10.1016/j.jmpt.2010.08.020.

Abstract

OBJECTIVE

Spinal manipulation therapy (SMT), an intervention used to treat low back pain, has been demonstrated to affect the stiffness of the spine. To adequately quantify the effects of SMT on stiffness, a device capable of applying specific parameters of manipulation in addition to measuring force-displacement values has been developed previously. Previously developed indentation techniques that quantify stiffness have been modified for novel use in evaluating SMT parameters. The reliability of stiffness measurements performed by the newly adapted device was assessed in this study.

METHODS

Seven springs of varying stiffness were each indented 10 times by a variable rate force/displacement (VRFD) device. Indentations were performed at a rate of 0.5 mm/s to a maximal displacement of 4 mm. The stiffness coefficients for a middle portion of the resulting force-displacement graph and the terminal instantaneous stiffness (stiffness at end displacement) were calculated. The intraclass correlation and confidence interval were calculated for these stiffness measurements to assess device reliability.

RESULTS

Repeated spring stiffness measures yielded an intraclass correlation coefficient value of 1.0. The mean stiffness values had narrow 95% confidence intervals ranging from 0.01 N/mm to 0.06 N/mm and small coefficients of variation.

CONCLUSION

This VRFD device provides highly reliable stiffness measurements in controlled conditions. Although in vivo reliability remains to be established, the results of this study support the use of the VRFD device in future trials investigating the impact of various SMT parameters on spinal stiffness.

摘要

目的

脊柱推拿疗法(SMT)是一种用于治疗腰痛的干预措施,已被证明会影响脊柱的刚度。为了充分量化SMT对刚度的影响,先前已开发出一种除了测量力-位移值之外还能够施加特定推拿参数的装置。先前开发的用于量化刚度的压痕技术已被修改,用于评估SMT参数的新用途。本研究评估了新改装装置进行的刚度测量的可靠性。

方法

使用可变速率力/位移(VRFD)装置对七个不同刚度的弹簧各进行10次压痕。压痕以0.5 mm/s的速率进行,最大位移为4 mm。计算所得力-位移图中间部分的刚度系数和末端瞬时刚度(末端位移时的刚度)。计算这些刚度测量值的组内相关性和置信区间,以评估装置的可靠性。

结果

重复的弹簧刚度测量得出组内相关系数值为1.0。平均刚度值的95%置信区间较窄,范围从0.01 N/mm到0.06 N/mm,变异系数较小。

结论

该VRFD装置在受控条件下提供高度可靠的刚度测量。尽管体内可靠性仍有待确定,但本研究结果支持在未来试验中使用VRFD装置来研究各种SMT参数对脊柱刚度的影响。

相似文献

1
Performance and reliability of a variable rate, force/displacement application system.
J Manipulative Physiol Ther. 2010 Oct;33(8):585-93. doi: 10.1016/j.jmpt.2010.08.020.
2
The reliability of a posterior-to-anterior spinal stiffness measuring system in a population of patients with low back pain.
J Manipulative Physiol Ther. 2007 Feb;30(2):116-23. doi: 10.1016/j.jmpt.2006.12.006.
4
The effect of duration and amplitude of spinal manipulative therapy (SMT) on spinal stiffness.
Man Ther. 2012 Dec;17(6):577-83. doi: 10.1016/j.math.2012.06.006. Epub 2012 Jul 17.
5
The effect of application site of spinal manipulative therapy (SMT) on spinal stiffness.
Spine J. 2015 Jun 1;15(6):1332-8. doi: 10.1016/j.spinee.2013.07.480. Epub 2013 Oct 17.
8
Measuring the posteroanterior stiffness of the cervical spine.
Man Ther. 2008 Dec;13(6):520-8. doi: 10.1016/j.math.2007.07.007. Epub 2007 Oct 1.
10

引用本文的文献

1
Systematic review on biomechanical effects of high-velocity, low amplitude spinal manipulation.
PLoS One. 2025 Jul 18;20(7):e0328048. doi: 10.1371/journal.pone.0328048. eCollection 2025.
5
Effects of unilateral facet fixation and facetectomy on muscle spindle responsiveness during simulated spinal manipulation in an animal model.
J Manipulative Physiol Ther. 2013 Nov-Dec;36(9):585-94. doi: 10.1016/j.jmpt.2013.08.007. Epub 2013 Oct 23.
6
The effect of application site of spinal manipulative therapy (SMT) on spinal stiffness.
Spine J. 2015 Jun 1;15(6):1332-8. doi: 10.1016/j.spinee.2013.07.480. Epub 2013 Oct 17.
9
The effect of duration and amplitude of spinal manipulative therapy (SMT) on spinal stiffness.
Man Ther. 2012 Dec;17(6):577-83. doi: 10.1016/j.math.2012.06.006. Epub 2012 Jul 17.

本文引用的文献

1
Load-displacement-time characteristics of the spine under posteroanterior mobilisation.
Aust J Physiother. 1992;38(2):115-23. doi: 10.1016/S0004-9514(14)60556-0.
2
The effect of abdominal stabilization contractions on posteroanterior spinal stiffness.
Spine (Phila Pa 1976). 2008 Mar 15;33(6):694-701. doi: 10.1097/BRS.0b013e318166e034.
3
Posteroanterior movements in tender and less tender locations of the cervical spine.
Man Ther. 2009 Feb;14(1):28-35. doi: 10.1016/j.math.2007.09.003. Epub 2007 Nov 26.
4
Determination of torque-limits for human and cat lumbar spine specimens during displacement-controlled physiological motions.
Spine J. 2009 Jan-Feb;9(1):77-86. doi: 10.1016/j.spinee.2007.07.391. Epub 2007 Nov 5.
5
Postero-anterior movements of the cervical spine: repeatability of force displacement curves.
Man Ther. 2008 Aug;13(4):341-8. doi: 10.1016/j.math.2007.03.005. Epub 2007 Oct 22.
7
The reliability of a posterior-to-anterior spinal stiffness measuring system in a population of patients with low back pain.
J Manipulative Physiol Ther. 2007 Feb;30(2):116-23. doi: 10.1016/j.jmpt.2006.12.006.
8
The accuracy of ultrasonic indentation in detecting simulated bone displacement: a comparison of three techniques.
J Manipulative Physiol Ther. 2006 Feb;29(2):126-33. doi: 10.1016/j.jmpt.2005.11.015.
9
Paraspinal muscle spindle responses to the duration of a spinal manipulation under force control.
J Manipulative Physiol Ther. 2006 Jan;29(1):22-31. doi: 10.1016/j.jmpt.2005.11.014.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验