Design of a multi-element TEPC for neutron monitoring.

作者信息

Waker A J, Lori J

机构信息

Faculty of Energy Systems and Nuclear Science, UOIT, 2000 Simcoe St. N., Oshawa, ON, Canada L1H 7K4.

出版信息

Radiat Prot Dosimetry. 2011 Feb;143(2-4):463-6. doi: 10.1093/rpd/ncq538. Epub 2010 Dec 24.

Abstract

Tissue-equivalent proportional counters (TEPCs) have long been considered suitable candidate instruments for more accurate neutron monitors in nuclear power plants. It has also been recognised that the production of truly light-weight devices based on TEPCs requires further effort directed towards increasing their sensitivity and decreasing their physical size. This paper deals with the construction of a multi-element TEPC (METEPC) designed to have the sensitivity of a 12.7-cm (5-in.) diameter spherical TEPC, but with approximately one-tenth of its physical size. Construction of the METEPC is achieved by machining 61 elongated cylindrical cavities in a single block of A150 TE plastic. Comparative measurements carried out in neutron fields with mean energies ranging from 34 to 354 keV demonstrate that the METEPC constructed does match the sensitivity of a 5-in. spherical TEPC and that microdosimetric lineal energy spectra measured with both detectors have the same features and show the same changes with neutron radiation quality.

摘要

组织等效正比计数器(TEPCs)长期以来一直被认为是核电站中更精确中子监测器的合适候选仪器。人们也认识到,基于TEPC制造真正的轻型设备需要进一步努力提高其灵敏度并减小其物理尺寸。本文论述了一种多元素TEPC(METEPC)的构造,该METEPC设计成具有直径为12.7厘米(5英寸)的球形TEPC的灵敏度,但物理尺寸约为其十分之一。通过在一块A150 TE塑料中加工61个细长的圆柱形腔来实现METEPC的构造。在平均能量范围为34至354 keV的中子场中进行的对比测量表明,所构建的METEPC确实与5英寸球形TEPC的灵敏度相匹配,并且用这两种探测器测量的微剂量线能量谱具有相同的特征,并且随着中子辐射质量的变化呈现相同的变化。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验