Pontifícia Universidade Católica do Rio Grande do Sul, Porto Alegre, RS, Brazil.
J Pediatr (Rio J). 2011 Jan-Feb;87(1):57-62. doi: 10.2223/JPED.2059. Epub 2011 Jan 19.
To compare blood pressure measurements in newborn infants using the flush method, pulse oximetry and oscillometry according to Doppler findings.
Noninvasive blood pressure measurements were made using three techniques (flush method, pulse oximetry and oscillometry) in three groups of newborns: 15 normal term infants, 16 stable preterm infants, and 14 critically ill infants. All measurements were video recorded, edited separately, coded and analyzed independently by three neonatologists.
Fifty-seven measurements were made using each method. The flush method and pulse oximetry had a better correlation with Doppler findings than oscillometry (correlation coefficients: 0.89, 0.85, 0.71; p < 0.01). The difference between measurement means, their standard deviations and the 95% confidence intervals when compared with Doppler findings were: -5.2 ± 7.9 (-21.1:10.7) mmHg for the flush method; 0.4 ± 8.9 (-17.5:18.2) mmHg for pulse oximetry; and 6.4 ± 16.1 (-25.8:8.6) mmHg for oscillometry. The flush method had a better agreement with Doppler findings for the diagnosis of hypotension than oximetry and oscillometry.
The flush method and pulse oximetry seem to be useful techniques to measure systolic blood pressure in newborn infants; oscillometry had the poorest agreement with Doppler findings to detect hypotension.
根据多普勒发现,比较冲洗法、脉搏血氧饱和度仪和振荡法测量新生儿血压的结果。
使用三种技术(冲洗法、脉搏血氧饱和度仪和振荡法)对三组新生儿进行无创血压测量:15 名正常足月新生儿、16 名稳定早产儿和 14 名危重新生儿。所有测量均进行视频记录,分别编辑,由三名新生儿科医生独立编码和分析。
使用每种方法进行了 57 次测量。冲洗法和脉搏血氧饱和度仪与多普勒发现的相关性优于振荡法(相关系数:0.89、0.85、0.71;p<0.01)。与多普勒发现相比,测量均值、标准差和 95%置信区间的差异为:冲洗法为-5.2±7.9(-21.1:10.7)mmHg;脉搏血氧饱和度仪为 0.4±8.9(-17.5:18.2)mmHg;振荡法为 6.4±16.1(-25.8:8.6)mmHg。与血氧饱和度仪和振荡法相比,冲洗法在诊断低血压方面与多普勒发现的一致性更好。
冲洗法和脉搏血氧饱和度仪似乎是测量新生儿收缩压的有用技术;振荡法与多普勒发现的一致性最差,无法检测低血压。