Department of Radiology, University of Palermo, Palermo, 90127, Italy.
Eur J Radiol. 2012 Apr;81(4):663-70. doi: 10.1016/j.ejrad.2011.01.015. Epub 2011 Feb 8.
To assess contrast-enhanced US (CEUS), computed tomography (CT) and magnetic resonance (MR) imaging findings and serial changes of the treated area at follow-up in patients who underwent liver resection using a bipolar radiofrequency electrosurgical device.
Imaging findings of 27 patients with resected hepatocellular carcinomas (HCCs) (n=24) and metastases (n=3) (mean size: 2.6cm), were retrospectively evaluated. Two readers assessed: the (a) presence, (b) thickness, (c) shape and (d) echogenicity (CEUS)/attenuation (CT)/signal intensity (MR imaging) at coagulated site and the (e) presence of residual tumor of the bipolar radiofrequency electrosurgical device resection margin.
Follow-up was performed with either CT (n=20) or MR imaging (n=7) and repeated in 16 patients with CT (n=7), MR imaging (n=4), or both techniques (n=5). Four patients also had a single CEUS examination. At first imaging follow-up a peripheral halo was depicted at resection site (100%). A fluid collection within the surgical area was found in 67% of patients. During the following imaging examinations a progressive involution of both findings was observed, respectively, in 81% and 62% of patients. Viable tissue was detected in three patients (11%).
After liver resection with bipolar radiofrequency electrosurgical device successfully ablated tumor is demonstrated at imaging by an unenhancing partial linear peripheral halo, in most cases, surrounding a fluid collection reducing in size during follow-up.
评估使用双极射频电外科设备行肝切除术患者的增强超声(CEUS)、计算机断层扫描(CT)和磁共振(MR)成像表现及随访时治疗区域的连续变化。
回顾性分析 27 例接受肝切除术的肝细胞癌(HCC)(n=24)和转移瘤(n=3)患者(平均大小:2.6cm)的影像学表现。两位读者评估了:(a)凝固部位的存在、(b)厚度、(c)形状和(d)回声强度(CEUS)/衰减(CT)/信号强度(MR 成像),以及(e)双极射频电外科设备切除边缘的残余肿瘤的存在。
20 例行 CT 随访,7 例行 MR 成像随访,16 例患者重复 CT(n=7)、MR 成像(n=4)或两种技术(n=5)检查。4 例患者还进行了单次 CEUS 检查。在首次影像学随访时,在切除部位均显示出周边晕环(100%)。67%的患者在手术区域发现了液体聚集。在随后的影像学检查中,分别有 81%和 62%的患者观察到这两种发现逐渐消退。3 例患者(11%)检测到存活组织。
使用双极射频电外科设备进行肝切除后,在影像学上成功消融肿瘤表现为未增强的部分线性外周晕环,在大多数情况下,周围环绕着一个在随访过程中逐渐缩小的液体聚集。