Baczynskyj L, Bronson G E
Physical and Analytical Chemistry Research, Upjohn Company, Kalamazoo, MI 49001.
Rapid Commun Mass Spectrom. 1990 Dec;4(12):533-5. doi: 10.1002/rcm.1290041213.
The electrospray mass spectra of several standard proteins were recorded and their molecular weights determined. These were compared to Literature values obtained by other laboratories. The agreement was quite good. Seven recombinant growth hormones were then investigated by this technique. The determined molecular weights were in agreement with the theoretical values. Differentiation between six of the seven analogs could be made on the basis of the molecular weight determinations. Two of these analogs differed in molecular weight by only 1 Da and the accuracy of the mass measurements was not sufficient for distinguishing these two homologous proteins. The influence of solvent on the mass accuracy determinations was studied. It appears that solvents form clusters with the proteins in the electrospray ionization method. These clusters broaden the peaks corresponding to the multiply charged ions of the proteins and make the determination of centroids more difficult.
记录了几种标准蛋白质的电喷雾质谱,并测定了它们的分子量。将这些结果与其他实验室获得的文献值进行了比较。吻合度相当好。然后用该技术研究了七种重组生长激素。测定的分子量与理论值相符。根据分子量测定结果,可以区分这七种类似物中的六种。其中两种类似物的分子量仅相差1道尔顿,质量测量的精度不足以区分这两种同源蛋白质。研究了溶剂对质量精度测定的影响。在电喷雾电离法中,溶剂似乎会与蛋白质形成簇。这些簇使对应于蛋白质多电荷离子的峰变宽,使得质心的测定更加困难。