Suppr超能文献

基于计算机断层扫描定位像的参考点定位与传统胶片及轴向计算机断层扫描的比较。

Comparison of computed tomography scout based reference point localization to conventional film and axial computed tomography.

作者信息

Jiang Lan, Templeton Alistair, Turian Julius, Kirk Michael, Zusag Thomas, Chu James C H

机构信息

Department of Radiation Oncology, Rush University Medical Center, Chicago, IL 60612, USA.

出版信息

Med Dosim. 2011 Winter;36(4):410-5. doi: 10.1016/j.meddos.2010.10.002. Epub 2011 Mar 11.

Abstract

Identification of source positions after implantation is an important step in brachytherapy planning. Reconstruction is traditionally performed from films taken by conventional simulators, but these are gradually being replaced in the clinic by computed tomography (CT) simulators. The present study explored the use of a scout image-based reconstruction algorithm that replaces the use of traditional film, while exhibiting low sensitivity to metal-induced artifacts that can appear in 3D CT methods. In addition, the accuracy of an in-house graphical software implementation of scout-based reconstruction was compared with seed location reconstructions for 2 phantoms by conventional simulator and CT measurements. One phantom was constructed using a planar fixed grid of 1.5-mm diameter ball bearings (BBs) with 40-mm spacing. The second was a Fletcher-Suit applicator embedded in Styrofoam (Dow Chemical Co., Midland, MI) with one 3.2-mm-diameter BB inserted into each of 6 surrounding holes. Conventional simulator, kilovoltage CT (kVCT), megavoltage CT, and scout-based methods were evaluated by their ability to calculate the distance between seeds (40 mm for the fixed grid, 30-120 mm in Fletcher-Suit). All methods were able to reconstruct the fixed grid distances with an average deviation of <1%. The worst single deviations (approximately 6%) were exhibited in the 2 volumetric CT methods. In the Fletcher-Suit phantom, the intermodality agreement was within approximately 3%, with the conventional sim measuring marginally larger distances, with kVCT the smallest. All of the established reconstruction methods exhibited similar abilities to detect the distances between BBs. The 3D CT-based methods, with lower axial resolution, showed more variation, particularly with the smaller BBs. With a software implementation, scout-based reconstruction is an appealing approach because it simplifies data acquisition over film-based reconstruction without requiring any specialized equipment and does not carry risk of misreads caused by artifacts.

摘要

植入后源位置的识别是近距离放射治疗计划中的重要一步。传统上,重建是根据传统模拟器拍摄的胶片进行的,但在临床中这些胶片正逐渐被计算机断层扫描(CT)模拟器所取代。本研究探索了一种基于定位像的重建算法的应用,该算法取代了传统胶片的使用,同时对三维CT方法中可能出现的金属诱导伪影表现出低敏感性。此外,通过传统模拟器和CT测量,将基于定位像重建的内部图形软件实现的准确性与2个模体的籽源位置重建进行了比较。一个模体是使用直径1.5毫米的滚珠轴承(BBs)的平面固定网格构建的,间距为40毫米。第二个是嵌入聚苯乙烯泡沫塑料(陶氏化学公司,密歇根州米德兰)的弗莱彻-休特施源器,在其周围6个孔中每个孔插入一个直径3.2毫米的BB。通过计算籽源之间距离的能力(固定网格为40毫米,弗莱彻-休特施源器中为30 - 120毫米)对传统模拟器、千伏CT(kVCT)、兆伏CT和基于定位像的方法进行了评估。所有方法都能够重建固定网格的距离,平均偏差<1%。两种容积CT方法中出现了最严重的单个偏差(约6%)。在弗莱彻-休特施源器模体中,不同模态之间的一致性在约3%以内,传统模拟器测量的距离略大,kVCT测量的距离最小。所有既定的重建方法在检测BBs之间距离方面表现出相似的能力。基于三维CT的方法轴向分辨率较低,表现出更多变化,尤其是对于较小的BBs。通过软件实现,基于定位像的重建是一种有吸引力的方法,因为它比基于胶片的重建简化了数据采集,无需任何专门设备,并且不存在由伪影导致误读的风险。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验