Safety, efficacy, and performance of implanted recycled cardiac rhythm management (CRM) devices in underprivileged patients.

作者信息

Hasan Reema, Ghanbari Hamid, Feldman Dustin, Menesses Daniel, Rivas Daniel, Zakhem Nicole C, Duarte Carlos, Machado Christian

机构信息

Division of Cardiac Electrophysiology and Cardiology, Providence Hospital and Medical Center, Wayne State University, Southfield, Michigan 48705, USA.

出版信息

Pacing Clin Electrophysiol. 2011 Jun;34(6):653-8. doi: 10.1111/j.1540-8159.2011.03061.x. Epub 2011 Mar 21.

Abstract

INTRODUCTION

Patients in underdeveloped nations have limited access to life-saving medical technology including cardiac rhythm management (CRM) devices. We evaluated alternative means to provide such technology to this patient population while assessing the safety and efficacy of such a practice.

METHODS

Patients in the United States with clinical indications for extraction of CRM devices were consented. Antemortem CRM devices were cleaned and sterilized following a protocol established at our institution. Surveillance in vitro cultures were performed for quality assurance. The functional status of pulse generators was tested with a pacing system analyzer to confirm at least 70% battery life. Most generators were transported, in person, to an implanting institution in Nicaragua. Recipients with a Class I indication for CRM implantation, and meeting economical criteria set forth, were consented for implantation of a recycled device.

RESULTS

Between 2003 and 2009, implantation was performed in 17 patients with an average age of 42.1 ± 20.3 years. Of the 17 patients, nine were male and eight were female. Mean follow-up was 68 ± 38 months. Device evaluation occurred prior to discharge, 4 weeks post implantation, and every 6 months thereafter. There were three deaths during the follow-up period secondary to myocardial infarction, stroke, and heart failure. Hematoma formation occurred in one patient. No infections, early battery depletion, or device malfunction were identified during follow-up.

CONCLUSIONS

Our case series is the longest follow-up of recipients of recycled antemortem CRM devices. Our findings support the feasibility and safety of this alternative acquisition of life-saving technology.