There are currently no comparison measurements of stress-induced changes in vascular function during acute mental stress tests to measurements made by BIOPAC MP150 systems technology, a standard polygraph device used to detect deception during polygraph examinations in military or law enforcement applications. Vascular responses to reactive hyperaemia and acute mental stress in 25 healthy subjects were measured by both peripheral arterial tonometry (EndoPAT) and a blood pressure cuff attached to a pressure transducer (BIOPAC) and compared. Reactive hyperaemia was performed at baseline and following three acute mental stress tests. There was no difference in vascular reactivity at baseline and following acute mental stress, as measured by EndoPAT or BIOPAC systems (p > 0·05). Mental stress ratios measured by EndoPAT were significantly different than those measured by BIOPAC (p < 0·01). These data suggest that EndoPAT measurements of vascular responses to acute mental stress may be more specific and sensitive than measurements using the BIOPAC system.
目前,还没有将急性心理应激测试期间血管功能的应激诱导变化的测量值与 BIOPAC MP150 系统技术(一种标准的测谎仪设备,用于在军事或执法应用中的测谎检查中检测欺骗)的测量值进行比较。通过外周动脉张力测量仪(EndoPAT)和连接到压力换能器的血压袖带(BIOPAC)测量了 25 名健康受试者的血管对反应性充血和急性心理应激的反应,并进行了比较。在基线和进行了三次急性心理应激测试后,分别进行了反应性充血。EndoPAT 或 BIOPAC 系统测量的基线和急性心理应激后的血管反应性没有差异(p>0·05)。EndoPAT 测量的心理应激比与 BIOPAC 测量的心理应激比显著不同(p<0·01)。这些数据表明,EndoPAT 对急性心理应激下血管反应的测量可能比使用 BIOPAC 系统的测量更具特异性和敏感性。