To evaluate the feasibility of using un-gated, real-time MRI for venous blood velocity mapping in the calf, comparing an interleaved spiral k-space sequence (ISP) against a standard segmented gradient echo sequence (GRE).
A flow phantom with a variable flow-rate was scanned using both GRE and ISP sequences for an in vitro comparison. Seven subjects were scanned prone, performing metronome guided breathing, using the (externally triggered) segmented GRE and real-time ISP sequences. The segmented GRE acquisition duration was 2.5 mins (22 guided respiratory cycles) and the ISP sequence ran continuously for 35 s, 4 full guided respiratory cycles. Mean velocity from each of the deep veins was measured and peak mean velocity, peak flow rate and cumulative volume flow over a respiratory cycle compared between sequences.
The two sequences compared well both in vitro and in vivo. The real-time ISP sequence showed short-term variations in mean velocity superimposed on the respiratory induced flow, which were averaged out using the segmented GRE sequence.
Real-time ISP provides comparable time-averaged flow results to the standard sequence with additional information on real-time flow variations and so could be used for further investigation into venous blood flow in the lower leg.
评估使用非门控实时 MRI 对小腿静脉血流速度进行映射的可行性,比较交错螺旋 k 空间序列(ISP)与标准分段梯度回波序列(GRE)。
使用 GRE 和 ISP 序列对具有可变流速的流动体模进行扫描,进行体外比较。七名受试者仰卧位,使用(外部触发)分段 GRE 和实时 ISP 序列进行节拍器引导呼吸。分段 GRE 采集时间为 2.5 分钟(22 个引导呼吸周期),而 ISP 序列连续运行 35 秒,4 个完整的引导呼吸周期。测量每条深静脉的平均速度,并比较序列之间的峰值平均速度、峰值流速和呼吸周期内的累积体积流量。
这两种序列在体外和体内都比较吻合。实时 ISP 序列显示呼吸引起的流动上叠加的短期平均速度变化,使用分段 GRE 序列平均化。
实时 ISP 序列提供了与标准序列相当的时间平均流速结果,并提供了实时流速变化的附加信息,因此可用于进一步研究小腿静脉血流。