Suppr超能文献

技术说明:用于妇科 HDR 近距离放疗的 3D 图像引导近距离放疗的锥形束 CT 成像。

Technical note: cone beam CT imaging for 3D image guided brachytherapy for gynecological HDR brachytherapy.

机构信息

Department of Radiation Oncology (MAASTRO), GROW Research Institute, University Medical Centre Maastricht, 6229ET Maastricht, The Netherlands.

出版信息

Med Phys. 2011 May;38(5):2762-7. doi: 10.1118/1.3578929.

Abstract

PURPOSE

This paper focuses on a novel image guidance technique for gynecological brachytherapy treatment. The present standard technique is orthogonal x-ray imaging to reconstruct the 3D position of the applicator when the availability of CT or MR is limited. Our purpose is to introduce 3D planning in the brachytherapy suite using a cone beam CT (CBCT) scanner dedicated to brachytherapy. This would avoid moving the patient between imaging and treatment procedures which may cause applicator motion. This could be used to replace the x-ray images or to verify the treatment position immediately prior to dose delivery.

METHODS

The sources of CBCT imaging artifacts in the case of brachytherapy were identified and removed where possible. The image quality was further improved by modifying the x-ray tube voltage, modifying the compensator bowtie filter and optimizing technical parameters such as the detector gain or tube current.

RESULTS

The image quality was adequate to reconstruct the applicators in the treatment planning system. The position of points A and the localization of the organs at risk (OAR) ICRU points is easily achieved. This allows identification of cases where the rectum had moved with respect to the ICRU point which would require asymmetrical source loading. A better visualization is a first step toward a better sparing of the OAR.

CONCLUSIONS

Treatment planning for gynecological brachytherapy is aided by CBCT images. CBCT presents advantages over CT: acquisition in the treatment room and in the treatment position due to the larger clearance of the CBCT, thereby reducing problems associated to moving patients between rooms.

摘要

目的

本文专注于一种新型的妇科近距离放射治疗图像引导技术。目前的标准技术是在 CT 或 MR 可用性有限的情况下,通过正交 X 射线成像来重建施源器的 3D 位置。我们的目的是在近距离治疗套房中使用专门用于近距离治疗的锥形束 CT(CBCT)扫描仪进行 3D 规划。这将避免在影像和治疗过程之间移动患者,因为这可能会导致施源器移动。这可以用于替代 X 射线图像,或者在剂量输送之前立即验证治疗位置。

方法

确定了在近距离治疗情况下 CBCT 成像伪影的来源,并尽可能去除。通过修改 X 射线管电压、修改补偿式射束挡块滤波器以及优化技术参数(如探测器增益或管电流),进一步提高了图像质量。

结果

图像质量足以在治疗计划系统中重建施源器。A 点的位置和危及器官(OAR)ICRU 点的定位很容易实现。这可以识别出直肠相对于 ICRU 点移动的情况,这种情况需要进行不对称的源加载。更好的可视化是更好地保护 OAR 的第一步。

结论

CBCT 图像有助于妇科近距离放射治疗的计划。CBCT 比 CT 具有优势:由于 CBCT 的更大间隙,可以在治疗室和治疗位置采集,从而减少了在房间之间移动患者的相关问题。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验