Suppr超能文献

使用在银纳米粒子沉积玻碳电极上的欠电位沉积来检测皮摩尔浓度范围内的铅离子。

Detection of lead ions in picomolar concentration range using underpotential deposition on silver nanoparticles-deposited glassy carbon electrodes.

机构信息

Department of Chemistry, Indian Institute of Technology Madras, Chennai 600 036, India.

出版信息

Talanta. 2011 Sep 30;85(4):2142-7. doi: 10.1016/j.talanta.2011.07.057. Epub 2011 Jul 23.

Abstract

The efficacy of silver-deposited glassy carbon electrode for the determination of lead ions at the sub-nanomolar concentration ranges is investigated. The silver nanoparticles are electrodeposited on glassy carbon electrode using chronoamperometry and the electrode surface is characterized using SEM. Lead ions are detected in the region of underpotential deposition. The analysis is performed in square wave mode in the stripping voltammetry without the removal of oxygen. The detection limit of 10 pM has been obtained with a constant potential of -0.7 V during the electrodeposition step for a period of 50s. The interference of surfactants in the detection of lead ions is also studied.

摘要

研究了沉积银的玻碳电极在亚纳摩尔浓度范围内测定铅离子的效果。采用计时安培法将银纳米粒子电沉积在玻碳电极上,并通过扫描电子显微镜对电极表面进行了表征。在欠电势沉积区域检测到铅离子。在不除氧的情况下,采用方波模式进行溶出伏安法分析。在电沉积步骤中,恒电位为-0.7V,持续 50s,检测限为 10pM。还研究了表面活性剂对铅离子检测的干扰。