Physiol Meas. 2011 Oct;32(10):1605-10. doi: 10.1088/0967-3334/32/10/008. Epub 2011 Sep 2.
An inflatable small plastic bag including a photo sensor was constructed for measurement of skin perfusion pressure avoiding the rim of the photo sensor over bony and tendineous surfaces of the tibia below the knee, at the ankle, and on the dorsal forefoot. Compression was obtained using a conical blood pressure cuff with continuous decrease from suprasystolic arm pressure. The validity of skin perfusion pressure with the new device was compared to that of isotope washout below the knee in normal subjects and in patients with an ischemic forefoot with acceptable agreement. The method had a high reproducibility within and between days in normal subjects. Compared to systolic arterial pressure measured using a strain gauge with a cuff on the ankle in normal subjects and patients with intermittent claudication the new device showed blood pressure in the skin closer to the diastolic pressure. The new pressure device thus had acceptable validity and reproducibility for estimation of the skin perfusion pressure and can be used on bony and tendineous sites on the lower limb in regions where critical wound healing is frequent, e.g. ankle and forefoot.