Department of Orthopedic Surgery, Nagoya University Graduate School of Medicine, Nagoya, Japan.
J Arthroplasty. 2012 May;27(5):820.e17-9. doi: 10.1016/j.arth.2011.08.018. Epub 2011 Sep 29.
A 61-year-old man who had undergone revision total hip arthroplasty in 2000 was evaluated at our clinic in 2008, 1 day after feeling an abnormal clicking sensation in his left hip. Radiographs showed a fracture of the alumina ceramic liner of his acetabular component. A second revision total hip arthroplasty was performed with a 28-mm highly crosslinked polyethylene acetabular liner and an alumina femoral head. Sixteen months after the second revision, the patient experience sudden-onset hip pain. Radiographs revealed a fatigue fracture of the neck of a displaced stem. Analysis by scanning electron microscopy of the surface of the retrieved alumina ceramic femoral head revealed no scratches or wear, but analysis of the highly crosslinked polyethylene socket revealed third-body wear, scratches, irregularly shaped dips, and white ceramic particles. Elemental analysis of the white particles by an x-ray microanalyzer revealed the presence of aluminum and oxygen.
一位 61 岁男性,2000 年曾行翻修全髋关节置换术,于 2008 年在我院就诊,1 天前出现左髋关节异常弹响声。X 线片显示髋臼假体的氧化铝陶瓷内衬骨折。行二次翻修全髋关节置换术,使用 28mm 高交联聚乙烯髋臼内衬和氧化铝股骨头。二次翻修后 16 个月,患者突发髋关节疼痛。X 线片显示脱位柄的颈部疲劳骨折。对取出的氧化铝陶瓷股骨头表面进行扫描电子显微镜分析显示无划痕或磨损,但对高交联聚乙烯髋臼分析显示有磨屑、划痕、形状不规则的凹陷和白色陶瓷颗粒。X 射线微分析仪对白色颗粒的元素分析显示存在铝和氧。