Department of Prosthodontics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, South Korea.
Int J Oral Maxillofac Implants. 2011 Sep-Oct;26(5):1016-23.
This study aimed to evaluate the effect of different screw-tightening sequences, torques, and methods on the strains generated on an internal-connection implant (Astra Tech) superstructure with good fit.
An edentulous mandibular master model and a metal framework directly connected to four parallel implants with a passive fit to each other were fabricated. Six stone casts were made from a dental stone master model by a splinted impression technique to represent a well-fitting situation with the metal framework. Strains generated by four screw-tightening sequences (1-2-3-4, 4-3-2-1, 2-4-3-1, and 2-3-1-4), two torques (10 and 20 Ncm), and two methods (one-step and two-step) were evaluated. In the two-step method, screws were tightened to the initial torque (10 Ncm) in a predetermined screw-tightening sequence and then to the final torque (20 Ncm) in the same sequence. Strains were recorded twice by three strain gauges attached to the framework (superior face midway between abutments). Deformation data were analyzed using multiple analysis of variance at a .05 level of statistical significance.
In all stone casts, strains were produced by connection of the superstructure, regardless of screw-tightening sequence, torque, and method. No statistically significant differences in superstructure strains were found based on screw-tightening sequences (range, -409.8 to -413.8 μm/m), torques (-409.7 and -399.1 μm/m), or methods (-399.1 and -410.3 μm/m).
Within the limitations of this in vitro study, screw-tightening sequence, torque, and method were not critical factors for the strain generated on a well-fitting internal-connection implant superstructure by the splinted impression technique. Further studies are needed to evaluate the effect of screw-tightening techniques on the preload stress in various different clinical situations.
本研究旨在评估不同的螺钉拧紧顺序、扭矩和方法对内连接种植体(AstraTech)上部结构在具有良好贴合度的情况下产生的应变的影响。
制作了一个无牙下颌主模型和一个金属框架,该框架直接连接到四个彼此具有被动配合的平行种植体上。通过夹板印模技术从牙科石工主模型制作了六个石铸模型,以代表金属框架具有良好贴合度的情况。评估了四种螺钉拧紧顺序(1-2-3-4、4-3-2-1、2-4-3-1 和 2-3-1-4)、两种扭矩(10 和 20 Ncm)和两种方法(一步法和两步法)产生的应变。在两步法中,螺钉按照预定的螺钉拧紧顺序拧紧至初始扭矩(10 Ncm),然后按照相同的顺序拧紧至最终扭矩(20 Ncm)。通过连接到框架(基牙之间的上表面中点)上的三个应变计两次记录应变。使用方差分析对变形数据进行分析,统计显著性水平为.05。
在所有石铸模型中,无论螺钉拧紧顺序、扭矩和方法如何,连接上部结构都会产生应变。根据螺钉拧紧顺序(范围为-409.8 至-413.8 μm/m)、扭矩(-409.7 和-399.1 μm/m)或方法(-399.1 和-410.3 μm/m),在上部结构应变方面没有发现统计学上的显著差异。
在本体外研究的限制范围内,螺钉拧紧顺序、扭矩和方法不是夹板印模技术对内连接种植体上部结构产生的应变的关键因素。需要进一步的研究来评估螺钉拧紧技术在各种不同临床情况下对预紧力的影响。