Hi-flexion and gender-specific designs fail to provide significant increases in range of motion during cruciate-retaining total knee arthroplasty.

机构信息

Center for Joint Disease, Chonnam National University Hwasun Hospital, Jeonnam, South Korea.

出版信息

J Arthroplasty. 2012 Jun;27(6):1081-4. doi: 10.1016/j.arth.2011.10.011. Epub 2011 Nov 23.

Abstract

The effects of different femoral component designs on intraoperative range of motion were examined in 40 female patients during primary cruciate-retaining (CR) total knee arthroplasty. After complete bone resection and soft tissue balancing, standard CR, high-flexion, and gender-specific knee trials were sequentially inserted, and maximal flexion and extension under gravity were measured using a navigation system. Average maximal flexions were 134.3° for standard CR knees, 136.2° for high-flexion knees, and 136.4° for gender-specific knees. No significant intergroup differences in intraoperative maximal flexion and extension were found (P > .05). High-flexion and gender-specific femoral designs were found to show subtle increases in intraoperative range of motion as compared with the standard design but no significant differences.

摘要

在 40 例初次行保留交叉韧带(CR)全膝关节置换术的女性患者中,研究了不同股骨组件设计对术中活动范围的影响。在完成骨切除和软组织平衡后,依次插入标准 CR、高屈曲和性别特异性膝关节试模,并使用导航系统测量重力下的最大屈曲和伸展。标准 CR 膝关节的平均最大屈曲度为 134.3°,高屈曲膝关节为 136.2°,性别特异性膝关节为 136.4°。术中最大屈曲和伸展的组间差异无统计学意义(P>.05)。与标准设计相比,高屈曲和性别特异性股骨设计在术中活动范围上略有增加,但无统计学差异。