Department of Radiology, Academic Medical Centre, Amsterdam, The Netherlands.
Magn Reson Med. 2012 Mar;67(3):595-600. doi: 10.1002/mrm.23215. Epub 2011 Dec 21.
Magnetic resonance imaging is widely used in the work-up and monitoring of patients with Crohn's disease. Balanced steady-state free precession sequences are an important part of the imaging protocol and until now primarily 1.5T scanners have been used in daily clinical practice. This is largely because running balanced steady-state free precession sequences in 3T magnets has technical problems related to increased B(0) inhomogeneity and specific absorption rate (SAR) deposition. A modified form of alternating repetition time steady-state free precession sequence is presented to acquire 3D-isotropic abdominal images with fat-suppression at 3T within a breath-hold. The modifications include an adjusted radiofrequency pulse shape, suitable phase-cycling scheme and TR(1)/TR(2) ratio. Results show that the proposed sequence is successful in obtaining high contrast 3D-isotropic abdominal images within a breath-hold. Furthermore, the proposed methodology is easy to implement in a clinical setting and does not require any postprocessing steps.
磁共振成像是克罗恩病患者的检查和监测的重要手段。平衡稳态自由进动序列是成像方案的重要组成部分,直到现在,1.5T 扫描仪在日常临床实践中主要用于平衡稳态自由进动序列。这主要是因为在 3T 磁体中运行平衡稳态自由进动序列存在与增加的 B(0)不均匀性和特定吸收率 (SAR)沉积相关的技术问题。提出了一种交替重复时间稳态自由进动序列的改进形式,以便在 3T 下在一次屏气中获得具有脂肪抑制的 3D 各向同性腹部图像。修改包括调整射频脉冲形状、合适的相循环方案和 TR(1)/TR(2) 比。结果表明,所提出的序列成功地在一次屏气中获得了高对比度的 3D 各向同性腹部图像。此外,所提出的方法易于在临床环境中实施,并且不需要任何后处理步骤。