Kaatze U, Scholle F D
Drittes Physikalisches Institut, Universität Göttingen.
Biomed Tech (Berl). 1990 Sep;35(9):190-4. doi: 10.1515/bmte.1990.35.9.190.
An apparatus is described which in the frequency range from about 5 to 500 MHz allows the ultrasonic absorption coefficient of biological tissues to be measured with 1% accuracy. This apparatus is based on a pulse transmission method in which a computer-controlled mode of operation is combined with a RF substitution technique. Superheterodyne detection of the transmitted signal, multiple data recording and signal averaging result in a high sensitivity of the measuring method. The apparatus can be also used to analyze the ultrasonic pulse reflected by the sample.
描述了一种装置,该装置在约5至500MHz的频率范围内能够以1%的精度测量生物组织的超声吸收系数。该装置基于脉冲传输方法,其中计算机控制的操作模式与射频替代技术相结合。对发射信号进行超外差检测、多数据记录和信号平均,使得该测量方法具有高灵敏度。该装置还可用于分析样品反射的超声脉冲。