Central Laboratory Animal Facility, University Hospital Schleswig-Holstein, Campus Kiel, Kiel, Germany.
J Thorac Cardiovasc Surg. 2012 Jun;143(6):1417-21. doi: 10.1016/j.jtcvs.2011.11.001. Epub 2012 Jan 12.
Malpositioning is one of the major problems in transcatheter aortic valve implantation. To evaluate the influence of mechanical balloon inflation on aortic valve stent positioning, the expansion process and the impact on the valve leaflet's structure were investigated.
Custom-made stents were laser cut from a 22-mm diameter stainless steel tube and mounted with a glutaraldehyde-treated bovine pericardial valve. The valved stents were crimped onto a standard balloon catheter and expanded by inflation of the balloon with 2 bar for 3 seconds. Expansion was studied using a high-speed camera, and the histology of the pericardial tissue was analyzed.
The valved stents were fully expanded within 3 seconds. Balloon inflation was observed to be asymmetric starting proximally. At the beginning of expansion, the valved stents were pulled proximally. During further inflation, the stents slipped distally on the balloon and experienced a total displacement of 13.5 mm. Macroscopic examination showed severe imprinting of the stent struts into the pericardial tissue. Histology revealed disrupted tissue layers and collagen fibers.
Analysis of valved stent expansion showed a displacement of the stent on the catheter during balloon inflation. Therefore, precise placement of the valved stent cannot be accomplished. Histologic analysis of the expanded pericardial tissue revealed disruption of collagen fibers. Disruption of pericardial tissue structures due to balloon expansion may result in early functional valve failure.
错位是经导管主动脉瓣植入术的主要问题之一。为了评估机械球囊充气对主动脉瓣支架定位的影响,研究了扩张过程以及对瓣叶结构的影响。
定制支架由 22 毫米直径的不锈钢管激光切割而成,并安装了戊二醛处理的牛心包瓣。带瓣支架被卷曲在标准球囊导管上,并通过球囊充气 2 巴 3 秒进行扩张。使用高速摄像机研究扩张过程,并分析心包组织的组织学。
带瓣支架在 3 秒内完全扩张。观察到球囊充气从近端开始不对称。在扩张开始时,带瓣支架被拉向近端。在进一步充气过程中,支架在球囊上向远端滑动,总位移为 13.5 毫米。宏观检查显示支架支柱严重压入心包组织。组织学显示组织层和胶原纤维破裂。
带瓣支架扩张分析显示球囊充气过程中支架在导管上发生移位。因此,无法精确放置带瓣支架。扩张的心包组织的组织学分析显示胶原纤维破裂。球囊扩张导致心包组织结构破坏可能导致早期功能瓣膜失效。