[超声检测微钙化——新方法的潜力]

[Sonographic detection of microcalcifications - potential of new method].

作者信息

Fischer T, Grigoryev M, Bossenz S, Diekmann F, Bick U, Slowinski T, Thomas A

机构信息

Institut of Radiology and Ultrasound Research Laboratory, Charité - Universitätsmedizin Berlin.

出版信息

Ultraschall Med. 2012 Aug;33(4):357-65. doi: 10.1055/s-0031-1299128. Epub 2012 Feb 9.

Abstract

PURPOSE

Does the easier microcalcification detection (EMD) method enable sonographic visualization of microcalcifications in breast core biopsy specimens compared with mammography?

MATERIALS AND METHODS

In a prospective randomized study, 105 core biopsy specimens obtained with stereotactic guidance were examined by mammography and ultrasound. EMD is integrated in a high-end ultrasound system and uses three level settings (0 - 2 blue, 3 - 5 violet, and 6 - 8 black-and-white; 14 MHz). Detection of microcalcifications per core specimen was determined for ultrasound and mammography. EMD image quality was rated on a scale of 1 - 9. ANOVA and Sidak post-hoc testing, Pearson regression analysis (r), and Spearman rank correlation (rho) were performed. The intraclass correlation coefficient (ICC) was calculated, and an ROC analysis was conducted.

RESULTS

The blue color map 1 was assigned the highest mean score of 1.5 ± 0.7 (p< 0.05 compared with black-and-white and violet). There was good correlation between the two modalities (r= 0.708 and rho= 0.694) with detection of 3.5 ± 3.1 microcalcifications per specimen by ultrasound versus 4.3 ± 4.8 by mammography (p> 0.05). The ICC of 0.773 indicates little disagreement between the two modalities. ROC analysis showed mammography to be superior to ultrasound compared with histological detection of microcalcifications (AUC= 0.837 vs. AUC= 0.728).

CONCLUSION

Sonographic detection of microcalcifications in stereotactic biopsy specimens using the EMD method correlates well with digital mammography. Mammography is slightly superior.

摘要

目的

与乳腺X线摄影相比,更简易的微钙化检测(EMD)方法能否实现乳腺粗针活检标本中微钙化的超声可视化?

材料与方法

在一项前瞻性随机研究中,对105例在立体定向引导下获取的粗针活检标本进行了乳腺X线摄影和超声检查。EMD集成于高端超声系统中,采用三种级别设置(0 - 2级为蓝色,3 - 5级为紫色,6 - 8级为黑白;14兆赫)。确定了超声和乳腺X线摄影对每个粗针标本微钙化的检测情况。EMD图像质量按1 - 9分进行评分。进行了方差分析和Sidak事后检验、Pearson回归分析(r)以及Spearman等级相关分析(rho)。计算了组内相关系数(ICC),并进行了ROC分析。

结果

蓝色图谱1的平均得分最高,为1.5 ± 0.7(与黑白和紫色图谱相比,p < 0.05)。两种检查方式之间存在良好相关性(r = 0.708,rho = 0.694),超声检测每个标本微钙化为3.5 ± 3.1个,乳腺X线摄影为4.3 ± 4.8个(p > 0.05)。ICC为0.773表明两种检查方式之间分歧较小。ROC分析显示,与微钙化的组织学检测相比,乳腺X线摄影优于超声(AUC = 0.837对AUC = 0.728)。

结论

使用EMD方法对立体定向活检标本进行超声微钙化检测与数字乳腺X线摄影具有良好相关性。乳腺X线摄影略占优势。