Fischer T, Grigoryev M, Bossenz S, Diekmann F, Bick U, Slowinski T, Thomas A
Institut of Radiology and Ultrasound Research Laboratory, Charité - Universitätsmedizin Berlin.
Ultraschall Med. 2012 Aug;33(4):357-65. doi: 10.1055/s-0031-1299128. Epub 2012 Feb 9.
Does the easier microcalcification detection (EMD) method enable sonographic visualization of microcalcifications in breast core biopsy specimens compared with mammography?
In a prospective randomized study, 105 core biopsy specimens obtained with stereotactic guidance were examined by mammography and ultrasound. EMD is integrated in a high-end ultrasound system and uses three level settings (0 - 2 blue, 3 - 5 violet, and 6 - 8 black-and-white; 14 MHz). Detection of microcalcifications per core specimen was determined for ultrasound and mammography. EMD image quality was rated on a scale of 1 - 9. ANOVA and Sidak post-hoc testing, Pearson regression analysis (r), and Spearman rank correlation (rho) were performed. The intraclass correlation coefficient (ICC) was calculated, and an ROC analysis was conducted.
The blue color map 1 was assigned the highest mean score of 1.5 ± 0.7 (p< 0.05 compared with black-and-white and violet). There was good correlation between the two modalities (r= 0.708 and rho= 0.694) with detection of 3.5 ± 3.1 microcalcifications per specimen by ultrasound versus 4.3 ± 4.8 by mammography (p> 0.05). The ICC of 0.773 indicates little disagreement between the two modalities. ROC analysis showed mammography to be superior to ultrasound compared with histological detection of microcalcifications (AUC= 0.837 vs. AUC= 0.728).
Sonographic detection of microcalcifications in stereotactic biopsy specimens using the EMD method correlates well with digital mammography. Mammography is slightly superior.
与乳腺X线摄影相比,更简易的微钙化检测(EMD)方法能否实现乳腺粗针活检标本中微钙化的超声可视化?
在一项前瞻性随机研究中,对105例在立体定向引导下获取的粗针活检标本进行了乳腺X线摄影和超声检查。EMD集成于高端超声系统中,采用三种级别设置(0 - 2级为蓝色,3 - 5级为紫色,6 - 8级为黑白;14兆赫)。确定了超声和乳腺X线摄影对每个粗针标本微钙化的检测情况。EMD图像质量按1 - 9分进行评分。进行了方差分析和Sidak事后检验、Pearson回归分析(r)以及Spearman等级相关分析(rho)。计算了组内相关系数(ICC),并进行了ROC分析。
蓝色图谱1的平均得分最高,为1.5 ± 0.7(与黑白和紫色图谱相比,p < 0.05)。两种检查方式之间存在良好相关性(r = 0.708,rho = 0.694),超声检测每个标本微钙化为3.5 ± 3.1个,乳腺X线摄影为4.3 ± 4.8个(p > 0.05)。ICC为0.773表明两种检查方式之间分歧较小。ROC分析显示,与微钙化的组织学检测相比,乳腺X线摄影优于超声(AUC = 0.837对AUC = 0.728)。
使用EMD方法对立体定向活检标本进行超声微钙化检测与数字乳腺X线摄影具有良好相关性。乳腺X线摄影略占优势。