Leunens G, Van Dam J, Dutreix A, van der Schueren E
Department of Radiotherapy, University Hospital St. Rafaël, Leuven, Belgium.
Radiother Oncol. 1990 Sep;19(1):73-87. doi: 10.1016/0167-8140(90)90167-u.
Combined entrance and exit dose measurements were performed with semiconductor detectors on patients, treated for neck and oral cavity malignancies. Transmission measurements showed the important influence of contour inaccuracies and tissue inhomogeneities. In 39.6% (21/53) of the checked contours, the discrepancy between the contour diameter used for routine treatment planning and the actual patient diameter was 1 cm or more, and in this group a systematic tendency for patient diameter underestimation due to the procedure was detected. When the X-ray beam passed through important bone structures such as the mandibular bones or the vertebral body, large discrepancies of 10% and more between the measured and the expected transmission were found. The target absorbed dose was determined from the transmission and entrance dose measurement. A systematic underdosage of about 2% at midline level was found to be due to an inaccuracy in the algorithms of the treatment planning system. Underdosages of 5% or more at midline were detected in more than 20% (47/230) of the measurements. In all cases, the reason for erroneous dose delivery was identified. Entrance dose measurements were previously demonstrated to be useful for the assessment of uncertainties related to treatment machine, patient set-up and treatment planning system (part 1). Transmission measurements (the ratio of the exit to the entrance dose measurement) are shown to be very useful to evaluate uncertainties related to patient data such as contour errors and tissue inhomogeneities as well as to the algorithms of the planning system. The influence of these errors on the target absorbed dose can be estimated and corrections can be applied for each individual patient.
使用半导体探测器对颈部和口腔恶性肿瘤患者进行了出入射剂量联合测量。透射测量显示了轮廓不准确和组织不均匀性的重要影响。在39.6%(21/53)的检查轮廓中,用于常规治疗计划的轮廓直径与患者实际直径之间的差异为1厘米或更大,并且在该组中检测到由于该程序导致患者直径低估的系统趋势。当X射线束穿过下颌骨或椎体等重要骨骼结构时,测量的透射率与预期透射率之间发现了10%及以上的大差异。目标吸收剂量由透射率和入射剂量测量确定。发现中线水平约2%的系统剂量不足是由于治疗计划系统算法不准确所致。在超过20%(47/230)的测量中检测到中线处剂量不足5%或更多。在所有情况下,都确定了错误剂量输送的原因。先前已证明入射剂量测量对于评估与治疗机器、患者摆位和治疗计划系统相关的不确定性很有用(第1部分)。透射测量(出射剂量测量与入射剂量测量的比值)对于评估与患者数据相关的不确定性(如轮廓误差和组织不均匀性)以及计划系统的算法非常有用。可以估计这些误差对目标吸收剂量的影响,并可以对每个患者进行校正。