On the behavior of instantaneous frequency estimators implemented on Doppler flow imagers.

作者信息

Nowicki A, Reid J, Pedersen P C, Schmidt A W, Oung H

机构信息

Institute of Fundamental Technological Research PAN, Warsaw, Poland.

出版信息

Ultrasound Med Biol. 1990;16(5):511-8. doi: 10.1016/0301-5629(90)90173-a.

Abstract

Analytical and experimental results have been used to examine the behavior of the "autocorrelator" or instantaneous frequency detector (IFD) applied to color-coded Doppler flow mapping. Two effects were studied. The first was the influence of noise, as modified by a stationary echo canceler, on the Doppler frequency detector. Our theoretical considerations predict that uncorrelated input noise signals become partially correlated after cancellation, and bias the response to flow signals. This effect was confirmed by experiment. The canceler introduces a constant negative bias into the denominator of the algorithm implemented by the estimator, thus changing the indicated frequency. The second phenomenon, examined through processing computer simulated Doppler signals added to real noise, is related to the possible ambiguity, called aliasing, of measurements of the mean frequency for wide-band Doppler spectra. We show that aliasing cannot be observed with these spectra unless the signal is first processed by a canceler. Thus, regions of apparent reversed flow direction on two-dimensional flow images of turbulence must usually be due to real reversal of the flow direction.

摘要

分析和实验结果已用于研究应用于彩色编码多普勒血流图的“自相关器”或瞬时频率检测器(IFD)的性能。研究了两种效应。第一种是由固定回波抵消器修正的噪声对多普勒频率检测器的影响。我们的理论分析预测,不相关的输入噪声信号在抵消后会部分相关,并使对血流信号的响应产生偏差。实验证实了这种效应。抵消器在估计器实现的算法分母中引入了恒定的负偏差,从而改变了指示频率。通过处理添加到真实噪声中的计算机模拟多普勒信号所研究的第二种现象,与宽带多普勒频谱平均频率测量中可能存在的模糊性(称为混叠)有关。我们表明,除非信号首先由抵消器处理,否则这些频谱中不会观察到混叠。因此,湍流二维血流图像上明显的血流方向反转区域通常必定是由于血流方向的实际反转。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验