Suppr超能文献

EVAHEART心室辅助泵心室抽吸条件的体外血流动力学评估

In vitro hemodynamic evaluation of ventricular suction conditions of the EVAHEART ventricular assist pump.

作者信息

Ferreira Antonio L, Yu Yih-Choung, Arnold Dorian K, Vandenberghe Stijn, Grashow Jonathan, Kitano Tomoya, Borzelleca David, Antaki James F

机构信息

Department of Biomedical Engineering, Carnegie Mellon University, Pittsburgh, Pennsylvania - USA and Department of Mathematics, Federal University of Maranhão, Maranhão - Brazil.

出版信息

Int J Artif Organs. 2012 Apr 30;35(4):263-71. doi: 10.5301/ijao.5000066. Epub 2012 Apr 11.

Abstract

PURPOSE

Mismatches between pump output and venous return in a continuous-flow ventricular assist device may elicit episodes of ventricular suction. This research describes a series of in vitro experiments to characterize the operating conditions under which the EVAHEART centrifugal blood pump (Sun Medical Technology Research Corp., Nagano, Japan) can be operated with minimal concern regarding left ventricular (LV) suction.

METHODS

The pump was interposed into a pneumatically driven pulsatile mock circulatory system (MCS) in the ventricular apex to aorta configuration. Under varying conditions of preload, afterload, and systolic pressure, the speed of the pump was increased step-wise until suction was observed. Identification of suction was based on pump inlet pressure.

RESULTS

In the case of reduced LV systolic pressure, reduced preload (=10 mmHg), and afterload (=60 mmHg), suction was observed for speeds=2,200 rpm. However, suction did not occur at any speed (up to a maximum speed of 2,400 rpm) when preload was kept within 10-14 mmHg and afterload=80 mmHg. Although in vitro experiments cannot replace in vivo models, the results indicated that ventricular suction can be avoided if sufficient preload and afterload are maintained.

CONCLUSION

Conditions of hypovolemia and/or hypotension may increase the risk of suction at the highest speeds, irrespective of the native ventricular systolic pressure. However, in vitro guidelines are not directly transferrable to the clinical situation; therefore, patient-specific evaluation is recommended, which can be aided by ultrasonography at various points in the course of support.

摘要

目的

在连续流心室辅助装置中,泵输出与静脉回流之间的不匹配可能引发心室抽吸事件。本研究描述了一系列体外实验,以确定在何种操作条件下,EVAHEART离心血泵(日本长野太阳医疗技术研究公司)在运行时对左心室(LV)抽吸的担忧最小。

方法

将该泵插入心室尖至主动脉配置的气动驱动脉动模拟循环系统(MCS)中。在不同的前负荷、后负荷和收缩压条件下,逐步提高泵的转速,直到观察到抽吸现象。根据泵入口压力来识别抽吸。

结果

在左心室收缩压降低、前负荷(=10 mmHg)和后负荷(=60 mmHg)降低的情况下,当转速=2200 rpm时观察到抽吸现象。然而,当前负荷保持在10 - 14 mmHg且后负荷=80 mmHg时,在任何转速(最高转速达2400 rpm)下均未出现抽吸现象。尽管体外实验不能替代体内模型,但结果表明,如果维持足够的前负荷和后负荷,可避免心室抽吸。

结论

低血容量和/或低血压状态可能会增加在最高转速下发生抽吸的风险,而与天然心室收缩压无关。然而,体外指导原则不能直接应用于临床情况;因此,建议进行针对患者的评估,在支持过程中的不同时间点可借助超声检查辅助评估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验