使用 Swan-Ganz 导管测定心输出量的改进。

Improvements in determination of cardiac output with a Swan-Ganz catheter.

作者信息

Sakagami M, Kuwana K, Nakanishi H, Sakai K

机构信息

Department of Chemical Engineering, Waseda University, Tokyo, Japan.

出版信息

ASAIO Trans. 1990 Jul-Sep;36(3):M684-6.

PMID:2252783
Abstract

The time constant for heat transfer may affect exact determination of cardiac output with Swan-Ganz catheters. Commercially available Swan-Ganz catheters are provided with thermistors with varying time constants. Current monitoring of cardiac output is not corrected for these time constants, so the conventional method of determining cardiac output using the equation of Stewart-Hamilton produces marked errors. The authors propose a new method of determining cardiac output with Swan-Ganz catheters with varying time constants from thermal dilution curve data based on Newton's cooling law. Values for blood flow rate determined by the new method using a completely stirred tank of original design, mimicking the natural heart and using bovine blood, are almost the same as values observed at varying saline infusion volumes, saline temperatures, and saline infusion times.

摘要

热传递的时间常数可能会影响使用 Swan-Ganz 导管精确测定心输出量。市售的 Swan-Ganz 导管配备有时间常数各异的热敏电阻。目前的心输出量监测并未针对这些时间常数进行校正,因此使用 Stewart-Hamilton 公式测定心输出量的传统方法会产生显著误差。作者提出了一种新方法,可根据牛顿冷却定律,利用热稀释曲线数据,对时间常数各异的 Swan-Ganz 导管测定心输出量。采用全新设计的完全搅拌槽模拟天然心脏并使用牛血,通过新方法测定的血流速率值,与在不同盐水输注量、盐水温度和盐水输注时间下观察到的值几乎相同。