Suppr超能文献

韩国人膝关节中当前全膝关节假体的适配度与术中人体测量结果的比较

The correctness of fit of current total knee prostheses compared with intra-operative anthropometric measurements in Korean knees.

作者信息

Ha C W, Na S E

机构信息

Samsung Medical Center, SungKyunKwan University School of Medicine, Department of Orthopaedic Surgery, KangNam Gu, IrWon Dong 50, Seoul 135-710, Korea.

出版信息

J Bone Joint Surg Br. 2012 May;94(5):638-41. doi: 10.1302/0301-620X.94B5.28824.

Abstract

We aimed to obtain anthropometric data on Korean knees and to compare these with data on commonly available total knee arthroplasties (TKAs). The dimensions of the femora and tibiae of 1168 knees were measured intra-operatively. The femoral components were found to show a tendency toward mediolateral (ML) under-coverage in small femurs and ML overhang in the large femurs. The ML under-coverage was most prominent for the small prostheses. The ML/anteroposterior (ML/AP) ratio of Korean tibiae was greater than that of tibial components. This study shows that, for different reasons, current TKAs do not provide a reasonable fit for small or large Korean knees, and that the 'gender-specific' and 'stature-specific' components help for large Korean femurs but offer less satisfactory fits for small femurs. Specific modifications of prostheses are needed for Asian knees.

摘要

我们旨在获取韩国人膝关节的人体测量数据,并将这些数据与常用全膝关节置换术(TKA)的数据进行比较。术中测量了1168个膝关节的股骨和胫骨尺寸。结果发现,股骨组件在小股骨中显示出向内外侧(ML)覆盖不足的趋势,而在大股骨中则表现为ML悬垂。ML覆盖不足在小假体中最为明显。韩国人胫骨的ML/前后径(ML/AP)比值大于胫骨组件的该比值。本研究表明,由于不同原因,目前的TKA对于大小不同的韩国人膝关节均不能提供合理的适配,“性别特异性”和“身高特异性”组件对大的韩国人股骨有帮助,但对小股骨的适配效果不太令人满意。亚洲人膝关节需要对假体进行特定修改。