Key Laboratory of Analytical Chemistry for Biology and Medicine (Ministry of Education), Department of Chemistry, Wuhan University, Wuhan 430072, PR China.
Talanta. 2012 May 30;94:70-6. doi: 10.1016/j.talanta.2012.02.053. Epub 2012 Mar 1.
A new method of solidified floating organic drop microextraction (SFODME) combined with electrothermal vaporization (ETV)-inductively coupled plasma mass spectrometry (ICP-MS) was developed for the determination of trace heavy metals in environmental water samples with sodium diethyldithiocarbamate (DDTC) as both chelating reagent in SFODME and chemical modifier in ETV. The factors affecting the microextraction efficiency were studied in detail and the optimal extraction conditions were established. Under the optimal conditions, the limits of detection (LODs) for SFODME-ETV-ICP-MS determination of Co, Pd, Cd, Hg, Pb and Bi were found to be 0.0060, 0.0091, 0.0020, 0.0041, 0.0170 and 0.0041 ng mL(-1), respectively, with the relative standard deviations (RSDs) of 2.8-10.0% (c=0.5 ng mL(-1), n=7). The developed method was successfully applied to the analysis of six target metals in Yangtze River and East Lake water samples with recoveries ranging from 77.7 to 119.1%. To validate the accuracy of the method, a certified reference material of Environmental Water (GSBZ50009-88) was analyzed and the determined values were in good agreement with the certified values.
一种新的固态浮动有机液滴微萃取(SFODME)方法与电热蒸发(ETV)-电感耦合等离子体质谱(ICP-MS)相结合,用于环境水样中痕量重金属的测定,其中二乙二硫代氨基甲酸钠(DDTC)既是 SFODME 中的螯合试剂,也是 ETV 中的化学修饰剂。详细研究了影响微萃取效率的因素,并确定了最佳萃取条件。在最佳条件下,SFODME-ETV-ICP-MS 测定 Co、Pd、Cd、Hg、Pb 和 Bi 的检出限(LODs)分别为 0.0060、0.0091、0.0020、0.0041、0.0170 和 0.0041 ng mL(-1),相对标准偏差(RSDs)为 2.8-10.0%(c=0.5 ng mL(-1),n=7)。该方法成功应用于长江和东湖水样中六种目标金属的分析,回收率在 77.7-119.1%之间。为了验证该方法的准确性,对环境水样的一种有证参考物质(GSBZ50009-88)进行了分析,测定值与证书值吻合良好。