J Magn Reson. 2012 Jun;219:41-5. doi: 10.1016/j.jmr.2012.03.007. Epub 2012 Apr 5.
Most 2D NMR spectra show artifacts that become increasingly more prominent as the relaxation delay between transients is decreased. Additionally, "rushing" a 2D experiment may lead to reduced sensitivity. It is shown here how to collect a DQF-COSY spectrum in less time, without artifacts, and with improved sensitivity, by a hardware solution we call Moving Tube NMR (MT NMR): the sample volume is physically moved out of the receiver coil after each transient and replaced by a fresh aliquot that is nearer to the equilibrium magnetization M(0). MT NMR was implemented with an automated mechanism that gave accurate and reproducible vertical tube movement, and a very long 5mm outer diameter (OD) NMR tube to hold a larger sample volume. Comparison of conventional and MT NMR DQF-COSY showed increased sensitivity and far reduced artifacts in the latter. The so-called t(1)-noise in the MT spectrum was no worse than in the conventional spectrum, pointing to the excellent specifications of the long 5mm OD tube, and the good mechanical handling of the automated drive. Thus, MT NMR could improve throughput for routine 2D NMR experiments without reducing sensitivity or adding artifacts, if sufficient sample is available. MT NMR could also be useful in cases of limited solubility, or for nuclei with long T(1) relaxation times.
大多数 2D NMR 光谱显示出伪影,随着瞬变之间的弛豫时间减少,伪影变得越来越明显。此外,“匆忙”进行 2D 实验可能会导致灵敏度降低。本文展示了如何通过我们称之为移动管 NMR(MT NMR)的硬件解决方案,在不产生伪影的情况下,以更少的时间收集 DQF-COSY 光谱,并提高灵敏度:在每次瞬变后,将样品体积从接收器线圈中移出,并替换为更接近平衡磁化强度 M(0)的新鲜等分试样。MT NMR 通过自动化机制实现,该机制可提供准确且可重复的垂直管运动,以及非常长的 5mm 外径(OD)NMR 管,可容纳更大的样品体积。比较常规和 MT NMR DQF-COSY 表明,后者的灵敏度更高,伪影大大减少。MT 光谱中的所谓 t(1)-噪声并不比常规光谱差,这表明长 5mm OD 管的优异规格和自动化驱动的良好机械处理。因此,如果有足够的样品,MT NMR 可以在不降低灵敏度或增加伪影的情况下提高常规 2D NMR 实验的通量。MT NMR 在溶解度有限的情况下或对于 T(1)弛豫时间长的核也可能有用。