Ennio Coscarelli Eye Clinic, Belo Horizonte, Brazil.
J Cataract Refract Surg. 2012 Jun;38(6):1006-13. doi: 10.1016/j.jcrs.2011.12.037.
To evaluate the clinical outcomes of implantation of Ferrara intrastromal corneal ring segments (ICRS) in patients with astigmatism after penetrating keratoplasty (PKP).
Private clinic, Belo Horizonte, Brazil.
Retrospective consecutive case series.
Chart records of post-PKP patients who had ICRS implantation from May 2005 to September 2009 were retrospectively reviewed. The following parameters were studied: corrected distance visual acuity (CDVA), keratometry (K) values, spherical equivalent (SE), spherical refractive error, corneal topographic astigmatism, minimum K, and maximum K.
The study evaluated 59 eyes (54 patients). The mean CDVA (logMAR) improved from 0.45 ± 0.17 (SD) (range 0.18 to 1.00) to 0.30 ± 0.17 (range 0.00 to 1.00). The mean SE was -6.34 ± 3.40 diopters (D) (range 0.37 to -16.50 D) preoperatively and -2.66 ± 2.52 D (range 0.87 to -10.50 D) postoperatively. The mean spherical refractive error decreased from -7.10 ± 3.07 D (range 2.15 to 16.68 D) preoperatively to -3.46 ± 2.05 D (range 0.88 to 10.79 D) postoperatively. No patient lost visual acuity. The mean corneal topographic astigmatism decreased from 3.37 ± 1.51 D preoperatively to 1.69 ± 1.04 D postoperatively. The mean maximum K value decreased from 48.09 ± 2.56 D to 44.17 ± 2.67 D and the mean minimum K value, from 44.90 ± 2.54 D to 42.46 ± 2.63 D. All changes were statistically significant (P<.0001).
Intrastromal corneal ring segments effectively reduced corneal cylinder in patients with astigmatism after PKP.
评估 Ferrara 基质内角膜环段(ICRS)在穿透性角膜移植(PKP)后散光患者中的临床效果。
巴西贝洛奥里藏特的私人诊所。
回顾性连续病例系列。
回顾性分析 2005 年 5 月至 2009 年 9 月间接受 ICRS 植入的 PKP 后患者的图表记录。研究了以下参数:矫正远视力(CDVA)、角膜曲率(K)值、等效球镜(SE)、球镜屈光不正、角膜地形图散光、最小 K 值和最大 K 值。
本研究共评估了 59 只眼(54 例患者)。CDVA(logMAR)从 0.45 ± 0.17(SD)(范围 0.18 至 1.00)提高到 0.30 ± 0.17(范围 0.00 至 1.00)。术前 SE 平均为-6.34 ± 3.40 屈光度(D)(范围 0.37 至-16.50 D),术后为-2.66 ± 2.52 D(范围 0.87 至-10.50 D)。术前平均球镜屈光不正从-7.10 ± 3.07 D(范围 2.15 至 16.68 D)降至-3.46 ± 2.05 D(范围 0.88 至 10.79 D)。没有患者视力丧失。角膜地形图散光平均从术前 3.37 ± 1.51 D 降至术后 1.69 ± 1.04 D。最大 K 值平均值从 48.09 ± 2.56 D 降至 44.17 ± 2.67 D,最小 K 值平均值从 44.90 ± 2.54 D 降至 42.46 ± 2.63 D。所有变化均具有统计学意义(P<.0001)。
基质内角膜环段可有效降低 PKP 后散光患者的角膜散光。