Arauzo A, Guerrero E, Urtizberea A, Stankiewicz J, Rillo C
Servicio de Medidas Físicas, Universidad de Zaragoza, Zaragoza 50009, Spain.
Rev Sci Instrum. 2012 Jun;83(6):066106. doi: 10.1063/1.4729380.
A sample holder design for high temperature measurements in a commercial MPMS SQUID magnetometer from Quantum Design is presented. It fulfills the requirements for the simultaneous use of the oven and reciprocating sample option (RSO) options, thus allowing sensitive magnetic measurements up to 800 K. Alternating current susceptibility can also be measured, since the holder does not induce any phase shift relative to the ac driven field. It is easily fabricated by twisting Constantan© wires into a braid nesting the sample inside. This design ensures that the sample be placed tightly into a tough holder with its orientation fixed, and prevents any sample displacement during the fast movements of the RSO transport, up to high temperatures.