Suppr超能文献

The resolution of aneuploid DNA stem lines by flow cytometry: limitations imposed by the coefficient of variation and the percentage of aneuploid nuclei.

作者信息

Cusick E L, Milton J I, Ewen S W

机构信息

Department of Surgery and Pathology, University of Aberdeen, Scotland.

出版信息

Anal Cell Pathol. 1990 Apr;2(3):139-48.

PMID:2275862
Abstract

Factors important in the resolution of cell sub-populations with differing DNA contents were investigated using an EPICS C flow cytometer. Software is available for the EPICS C which permits data from any two histograms to be superimposed or added together before display. Samples of fresh and archival thyroid tissue, stained with propidium iodide, were analysed on the flow cytometer and the peak channel number noted. The photomultiplier (PMT) voltage was increased and the sample analysed again producing a second histogram with a higher peak channel number. The two histograms were added together to simulate a cell suspension with two sub-populations with a different DNA content. By systematically altering the PMT voltage and the number of nuclei included in each analysis, it was possible to examine the importance of DNA index and the percentage of tumor cells with an aneuploid DNA content for both fresh and paraffin-embedded thyroid nuclei. The crucial importance of achieving a low coefficient of variation (CV) was demonstrated and consequently the reservations that pertain when archival material is studied, particularly in tumours where DNA aneuploidy is frequently expressed with a low DNA index.