German Red Cross Blood Service Baden-Württemberg-Hessen, Institute Ulm, Ulm, Germany.
Int J Immunogenet. 2013 Jun;40(3):171-7. doi: 10.1111/j.1744-313X.2012.01147.x. Epub 2012 Aug 23.
The Luminex xMAP system has become an important tool for HLA antibody screening and identification in sera of transplant patients. Recently, the Luminex single antigen bead assay was shown to be prone to an artefact, the so called prozone phenomenon: Sera with high titer HLA antibodies gave negative results when tested neat, but reacted strongly positive after 1:10 dilution. We also observed such a phenomenon and found that it was most likely caused by the complement component 1 (C1) by competitively displacing the detection antibodies. In this article we review the complement-mediated prozone effect and other mechanisms of interference with solid phase assays, and we discuss possible consequences for HLA antibody testing with the Luminex SAB assay.
Luminex xMAP 系统已成为移植患者血清中 HLA 抗体筛选和鉴定的重要工具。最近,Luminex 单抗原珠检测法容易出现一种假象,即所谓的前区现象:高滴度 HLA 抗体的血清直接检测时结果为阴性,但 1:10 稀释后反应强烈阳性。我们也观察到了这种现象,发现它很可能是由补体成分 1(C1)竞争性取代检测抗体引起的。本文综述了补体介导的前区效应和固相检测干扰的其他机制,并讨论了 Luminex SAB 检测法进行 HLA 抗体检测的可能后果。