Qu Mingliang, Ehman Eric, Fletcher Joel G, Huprich James E, Hara Amy K, Silva Alvin C, Farrugia Gianrico, Limburg Paul, McCollough Cynthia H
Department of Radiology, Mayo Clinic, Rochester, MN, USA.
J Comput Assist Tomogr. 2012 Sep-Oct;36(5):554-9. doi: 10.1097/RCT.0b013e3182606baf.
To develop dual-energy computed tomography methods for identification of hyperenhancing, hypoenhancing, and nonenhancing small-bowel pathologies.
Small-bowel phantoms simulating varying patient sizes and polyp types (hyperenhancing, hypoenhancing, and nonenhancing) contained bismuth suspension in the lumen. Dual-energy CT was performed at 80/140 kV and 100/140 kV. Computed tomographic number ratios (CT numbers at low/high kilovoltage) were calculated. Two radiologists evaluated polyp detection and conspicuity using bismuth-only, iodine-only, iodine-overlay, and mixed-kilovoltage displays.
Computed tomographic ratios for bismuth and iodine did not overlap. For hyperenhancing and nonenhancing polyps at 80/140 kV, iodine-overlay display yielded higher detection rate (96%, 94%) and conspicuity score (3.5, 3.1) than mixed-kilovoltage images (88%, 68%; 1.5, 2.7). Mixed-kV images performed slightly better for hypoenhancing polyps (92%, 3.4 vs. <80%, <2.9). Similar results were observed at 100/140kV.
Dual-energy CT and a bismuth-containing enteric contrast permitted simultaneous identification of hyperenhancing, hypoenhancing, and nonenhancing polyps over a range of patient sizes.
开发双能计算机断层扫描方法,用于识别小肠增强、低增强和无增强的病变。
模拟不同患者体型和息肉类型(增强、低增强和无增强)的小肠模型,其管腔内含有铋悬浮液。在80/140 kV和100/140 kV下进行双能CT扫描。计算计算机断层扫描数比(低/高千伏下的CT数)。两名放射科医生使用仅铋、仅碘、碘叠加和混合千伏显示来评估息肉的检测和清晰度。
铋和碘的计算机断层扫描比值不重叠。对于80/140 kV下的增强和无增强息肉,碘叠加显示的检测率(96%,94%)和清晰度评分(3.5,3.1)高于混合千伏图像(88%,68%;1.5,2.7)。混合千伏图像对低增强息肉的表现略好(92%,3.4对<80%,<2.9)。在100/140 kV下观察到类似结果。
双能CT和含铋肠道造影剂能够在一系列患者体型范围内同时识别增强、低增强和无增强的息肉。