Department of Oral and Maxillofacial Radiology, Graduate School, Tokyo Medical and Dental University, Tokyo, Japan.
Implant Dent. 2012 Dec;21(6):496-501. doi: 10.1097/ID.0b013e3182703387.
To evaluate whether the measurement accuracy of dental CT images is affected by the mandibular positioning and the pitch factor at CT scanning.
Three dry mandibles were scanned by 64-slice multidetector row CT (MDCT) and micro-CT. For MDCT scanning, 7 different mandibular positioning and 3 different pitch factors were applied. After reformatting dental CT images, the bone height was measured on these images. It was also measured on the corresponding micro-CT image, which was defined as the actual value. The difference of the measurement values between these 2 was defined as the measurement error.
There was no significant difference in the measurement errors due to either the mandibular positioning or the pitch factor.
The measurement accuracy of dental CT images obtained was not affected by either mandibular positioning or pitch factor at CT scanning.
评估在 CT 扫描时下颌定位和螺距因素是否会影响牙 CT 图像的测量精度。
使用 64 层多排螺旋 CT(MDCT)和微 CT 对 3 个干燥的下颌骨进行扫描。对于 MDCT 扫描,应用了 7 种不同的下颌定位和 3 种不同的螺距因素。在对牙 CT 图像进行重新格式化后,在上面对图像进行骨高度测量。同时,也在对应的微 CT 图像上进行测量,将其定义为实际值。这两个值之间的差值被定义为测量误差。
无论是下颌定位还是螺距因素,测量误差均无显著差异。
在 CT 扫描时,下颌定位或螺距因素均不会影响牙 CT 图像的测量精度。