Institute of Biomedical Engineering, National Taiwan University, No. 1 Sec. 4, Roosevelt Road, Taipei, 10617, Taiwan.
Clin Chim Acta. 2013 Jan 16;415:152-7. doi: 10.1016/j.cca.2012.10.033. Epub 2012 Oct 23.
Glucose oxidase biosensors are used in self-monitoring blood glucose concentrations. The capillary blood glucose quantitation requires a calibration curve. Due to the limitation in obtaining calibration curve from capillary blood, an alternate approach by using venous blood for neonatal measurement was investigated.
A signal correlation between oxygen saturated venous blood and capillary blood was derived. The hematocrit effect was studied for different glucose concentrations. The calibrated glucose strips were validated by neonatal intensive care unit (NICU) samples.
A simple equation, finger blood signal=1.39∗(oxygen saturated venous blood signal)-31.2 was derived. The rate of change in glucose concentration due to hematocrit effect was low in lower glucose concentration samples. The BeneCheck Glucose Strips were compared with Beckman Coulter analyzer by using 52 NICU samples. More than 95% of test results were within the variation of ±10 mg/dl of bias and ±15% of bias% when glucose concentration is <75 mg/dl and ≥75 mg/dl respectively.
The BeneCheck Glucose Strips can be accurately calibrated with venous blood. The hematocrit effect can also be predicted. Based on this study, BeneCheck Blood Glucose Monitoring System can be suitable for neonatal glucose measurement.
葡萄糖氧化酶生物传感器用于自我监测血糖浓度。毛细血管血糖定量需要校准曲线。由于从毛细血管获得校准曲线的限制,因此研究了使用静脉血进行新生儿测量的替代方法。
推导了氧饱和静脉血和毛细血管血之间的信号相关性。研究了不同葡萄糖浓度下的红细胞压积效应。通过新生儿重症监护病房(NICU)样本验证了校准后的血糖试纸。
得出了一个简单的方程式,指尖血信号=1.39∗(氧饱和静脉血信号)-31.2。在葡萄糖浓度较低的样本中,由于红细胞压积效应导致的葡萄糖浓度变化率较低。使用 52 个 NICU 样本将 BeneCheck 血糖试纸与贝克曼库尔特分析仪进行比较。当血糖浓度<75mg/dl 和≥75mg/dl 时,95%以上的测试结果在偏差±10mg/dl 和偏差%±15%的范围内。
BeneCheck 血糖试纸可以通过静脉血进行准确校准。也可以预测红细胞压积效应。基于这项研究,BeneCheck 血糖监测系统可以适用于新生儿血糖测量。