Pacing Clin Electrophysiol. 2013 Jan;36(1):109-12. doi: 10.1111/pace.12041. Epub 2012 Nov 4.

Abstract

BACKGROUND

Few studies have examined the prevalence of permanent pacemaker (PPM) malfunction among patients with a previously implanted pacemaker admitted to the hospital with syncope.

OBJECTIVE

This study sought to examine causes of syncope in patients with a previously implanted pacemaker admitted to our hospital with syncope.

METHODS

We retrospectively reviewed our hospital admission database for patients who had both keywords "syncope" and "pacemaker" as their diagnoses from January 1, 1995 until June 1, 2012. One hundred and sixty-two patients who were admitted to the hospital because of syncope and had a PPM implanted prior to the index syncopal episode were included. All patients had pacemakers interrogated during the admission. Two independent physicians examined the discharge summary of each patient and determined the cause of syncope in each case.

RESULTS

Of the 162 patients studied, eight (4.9%) were found to have pacemaker system malfunction as a cause of syncope. In 96 patients (59.2%), the cause of syncope could not be determined prior to hospital discharge. Among the identifiable causes of syncope, orthostatic hypotension was most prevalent (16%) followed by vasovagal (6%), severe aortic stenosis (4.3%), atrial arrhythmia (3.1%), acute and subacute infection (3.1%), and other less prevalent causes (3.1%).

CONCLUSION

In this study, PPM system malfunction was rarely a cause of syncope in patients admitted to the hospital with a previously implanted device.

摘要

背景

很少有研究调查过因晕厥入院且此前已植入永久起搏器(PPM)的患者中PPM故障的发生率。

目的

本研究旨在调查因晕厥入住我院且此前已植入起搏器的患者发生晕厥的原因。

方法

我们回顾性分析了我院1995年1月1日至2012年6月1日期间诊断关键词同时包含“晕厥”和“起搏器”的患者入院数据库。纳入162例因晕厥入院且在此次晕厥发作前已植入PPM的患者。所有患者在入院期间均接受了起搏器问询。两名独立的医生检查了每位患者的出院小结,并确定了每例患者晕厥的原因。

结果

在研究的162例患者中,发现8例(4.9%)晕厥原因是起搏器系统故障。96例患者(59.2%)在出院前无法确定晕厥原因。在可明确的晕厥原因中,体位性低血压最为常见(16%),其次是血管迷走性晕厥(6%)、严重主动脉瓣狭窄(4.3%)、房性心律失常(3.1%)、急慢性感染(3.1%)以及其他不太常见的原因(3.1%)。

结论

在本研究中,对于因此前已植入装置而入院的患者,PPM系统故障很少是晕厥的原因。