Suppr超能文献

Absolute sensitivity calibration of vacuum and extreme ultraviolet spectrometer systems and Z(eff) measurement based on bremsstrahlung continuum in HL-2A tokamak.

作者信息

Zhou Hangyu, Cui Zhengying, Morita Shigeru, Fu Bingzhong, Goto Motoshi, Sun Ping, Dong Chunfeng, Gao Yadong, Xu Yuan, Lu Ping, Yang Qingwei, Duan Xuru

机构信息

Southwestern Institute of Physics, PO Box 432, Chengdu 610041, China.

出版信息

Rev Sci Instrum. 2012 Oct;83(10):10D507. doi: 10.1063/1.4729671.

Abstract

A grazing-incidence flat-field extreme ultraviolet (EUV) spectrometer has been newly developed in HL-2A tokamak. Typical spectral lines are observed from intrinsic impurities of carbon, oxygen, iron, and extrinsic impurity of helium in the wavelength range of 20 Å-500 Å. Bremsstrahlung continuum is measured at different electron densities of HL-2A discharges to calibrate absolute sensitivity of the EUV spectrometer system and to measure effective ionic charge, Z(eff). The sensitivity of a vacuum ultraviolet (VUV) spectrometer system is also absolutely calibrated in overlapped wavelength range of 300 Å-500 Å by comparing the intensity between VUV and EUV line emissions.