Utilizing confocal microscopy to measure refractive index of articular cartilage.

机构信息

School of Mechanical and Chemical Engineering, University of Western Australia, Perth, WA, Australia.

出版信息

J Microsc. 2012 Dec;248(3):281-91. doi: 10.1111/j.1365-2818.2012.03674.x.

Abstract

This study proposes a method for measuring the refractive index of articular cartilage within a thin and small specimen slice. The cartilage specimen, with a thickness of about 50 μm, was put next to a thin film of immersion oil of similar thickness. Both the articular cartilage and immersion oil were scanned along the depth direction using a confocal microscope. The refractive index mismatch between the cartilage and the immersion oil induced a slight axial deformation in the confocal images of the cartilage specimen that was accurately measured by a subpixel edge-detection-based technique. A theoretical model was built to quantify the focal shift of confocal microscopy caused by the refractive index mismatch. With the quantitative deformations of cartilage images and the quantified function of focal shift, the refractive index of articular cartilage was accurately interpolated. At 561 nm, 0.1 MPa and 20 °C, the overall refractive index of the six cartilage plugs was 1.3975 ± 0.0156. The overall coefficient of variation of all cartilage specimens was 0.68%, which indicated the high repeatability of our method. The verification experiments using distilled water showed a minimal relative error of 0.02%.

摘要

本研究提出了一种测量薄小软骨样本折射率的方法。将厚度约为 50μm 的软骨样本放置在厚度相似的薄浸油膜旁边。使用共焦显微镜沿深度方向扫描关节软骨和浸油。折射率不匹配会导致软骨样本共焦图像产生微小的轴向变形,通过基于亚像素边缘检测的技术可以准确测量这种变形。建立了一个理论模型来量化折射率不匹配引起的共焦显微镜的焦移。通过定量的软骨图像变形和定量的焦移函数,可以准确地插值出关节软骨的折射率。在 561nm、0.1MPa 和 20°C 下,六个软骨塞的整体折射率为 1.3975±0.0156。所有软骨样本的整体变异系数为 0.68%,表明我们的方法具有很高的重复性。使用蒸馏水进行的验证实验显示相对误差最小为 0.02%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验