The validity of different display monitors in the assessment of dental implant site dimensions in cone beam computed tomography images.

机构信息

Department of Oral Medicine and Diagnostic Sciences, College of Dentistry, King Saud University, Riyadh, Saudi Arabia.

出版信息

Acta Odontol Scand. 2013 Sep;71(5):1085-91. doi: 10.3109/00016357.2012.741709. Epub 2012 Nov 21.

Abstract

OBJECTIVE

To determine if the differing contrast resolution of various LCD monitors affects the reliability or accuracy of measurements of proposed implant sites.

MATERIALS AND METHODS

Edentulous areas of human dry skulls were marked with radiopaque markers in order to standardize the plane of the transverse cross-sections of the ridges and path of measurements. The skulls were imaged by a CBCT device and the images stored in proprietary format on the workstation. The data sets were then transferred, with the proprietary reformatting software, to two different laptops using CDs. Transverse cross-sectional images of the ridges were reformatted on all three computers and ridge dimensions were recorded using the linear measurement tool of the proprietary software. Ridge dimensions were recorded directly from the three different monitors by two observers and compared to measurements recorded directly from the bone. The measurement errors and intra- and inter-examiner reliability were calculated for each monitor and compared with each other.

RESULTS

Intra- and inter-examiner reliability scores for the measurements recorded from all three devices were very high and ranged between 0.993-0.999. The mean of the absolute errors was 0.55 mm for the workstation, 0.61 mm for laptop 1 and 0.68 mm for laptop 2. The absolute errors were statistically significant for all three monitors (p-value < 0.001), but there was no statistically significant difference between the absolute errors obtained from the three monitors.

CONCLUSIONS

No differences in the reliability or accuracy of measurements of implant site dimensions were obtained using color LCD monitors with different contrast resolution capabilities.

摘要

目的

确定不同液晶显示器的对比度分辨率是否会影响拟议种植体部位测量的可靠性或准确性。

材料与方法

用人颅骨的干燥标本,在其上用不透射线的标记物标记,以标准化牙槽嵴的横断面平面和测量路径。使用 CBCT 设备对颅骨进行成像,并将图像以专有格式存储在工作站上。然后使用专有格式转换软件将数据集通过 CD 传输到两台不同的笔记本电脑上。在所有三台计算机上对牙槽嵴的横断面图像进行重新格式化,并使用专有软件的线性测量工具记录牙槽嵴尺寸。两名观察者直接从三个不同的监视器上记录牙槽嵴尺寸,并与直接从骨头上记录的测量值进行比较。计算了每个监视器的测量误差和观察者内、观察者间可靠性,并将其相互比较。

结果

从所有三个设备记录的测量值的观察者内和观察者间可靠性评分均非常高,范围在 0.993-0.999 之间。工作站的平均绝对误差为 0.55 毫米,笔记本电脑 1 为 0.61 毫米,笔记本电脑 2 为 0.68 毫米。三个监视器的绝对误差均具有统计学意义(p 值<0.001),但三个监视器的绝对误差之间无统计学差异。

结论

使用对比度分辨率不同的彩色液晶显示器,在种植体部位尺寸的测量可靠性或准确性方面没有差异。