National Tuberculosis Reference Laboratory, National Center for Tuberculosis Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing, People's Republic of China.
Int J Tuberc Lung Dis. 2013 Jan;17(1):107-12. doi: 10.5588/ijtld.12.0184.
To assess the feasibility of using light-emitting diode fluorescence microscopy (LED-FM) in peripheral laboratories in China.
The performance of LED-FM and Ziehl-Neelsen (ZN) microscopy was compared on slides directly prepared from the sputum of tuberculosis (TB) suspects and follow-up patients on treatment. The examination time, fading of fluorescence-stained slides, average unit cost and qualitative user appraisal of LED-FM were also analysed.
Among 11 276 slides, the smear-positive rate for LED-FM was 11.2% (1263/11 276), 2.6% (294/11 276) higher than that of ZN (8.6%, 969/11 276; χ(2) 263.5, P < 0.05). The examination time for LED-FM (120.0 ± 38.9 seconds) was shorter than that for ZN (206.3 ± 75.9 s; t = 28.12, P < 0.05). For smear fading, quantitative and qualitative errors occurred within respectively 7.8 and 7.7 weeks. The average unit costs for ZN and LED-FM were respectively US$2.20 ± 0.58 and US$1.97 ± 0.71 (t = 5.08, P < 0.05). LED-FM was accepted by most laboratory technicians.
LED-FM compared favourably with ZN, with a higher smear-positive detection rate, a shorter examination time and lower unit examination cost. LED-FM may be an alternative to ZN as a cost-effective method for detecting bacilli in peripheral laboratories in China.
评估在中国的基层实验室使用发光二极管荧光显微镜(LED-FM)的可行性。
对直接取自肺结核(TB)疑似患者和治疗随访患者痰液标本的载玻片,比较 LED-FM 和齐-尼(ZN)显微镜的性能。还分析了 LED-FM 的检查时间、荧光染色载玻片褪色、平均单位成本和定性用户评估。
在 11276 张载玻片标本中,LED-FM 的阳性检出率为 11.2%(1263/11276),比 ZN 高 2.6%(8.6%,969/11276;χ(2)263.5,P<0.05)。LED-FM 的检查时间(120.0±38.9 秒)短于 ZN(206.3±75.9 s;t=28.12,P<0.05)。对于涂片褪色,定量和定性误差分别在 7.8 和 7.7 周内发生。ZN 和 LED-FM 的平均单位成本分别为 2.20±0.58 美元和 1.97±0.71 美元(t=5.08,P<0.05)。大多数实验室技术人员都接受了 LED-FM。
LED-FM 与 ZN 相比具有优势,阳性检出率更高,检查时间更短,单位检查成本更低。LED-FM 可能是 ZN 的替代方法,是中国基层实验室检测杆菌的一种具有成本效益的方法。