The copper intrauterine device (IUD) based its contraceptive action on the release of cupric ions from a copper wire. Immediately after the insertion, a burst release of copper ions occurs, which may be associated to a variety of side effects. 6-Mercaptopurine (6-MP) and pterin (PT) have been proposed as corrosion inhibitors to reduce this harmful release. Pretreatments with 1 × 10(-4) M 6-MP and 1 × 10(-4) M PT solutions with 1h and 3h immersion times were tested. Conventional electrochemical techniques, EDX and XPS analysis, and cytotoxicity assays with HeLa cell line were employed to investigate the corrosion behavior and biocompatibility of copper with and without treatments. Results showed that copper samples treated with PT and 6-MP solutions for 3 and 1 h, respectively, are more biocompatible than those without treatment. Besides, the treatment reduces the burst release effect of copper in simulated uterine solutions during the first week after the insertion. It was concluded that PT and 6-MP treatments are promising strategies able to reduce the side effects related to the "burst release" of copper-based IUD without altering the contraceptive action.
基于铜丝释放铜离子的铜宫内节育器(IUD)具有避孕作用。放置后即刻会发生铜离子的爆发式释放,这可能与多种副作用有关。6-巯基嘌呤(6-MP)和蝶呤(PT)已被提议作为腐蚀抑制剂来减少这种有害的释放。本研究测试了 1×10(-4) M 6-MP 和 1×10(-4) M PT 溶液浸泡 1 h 和 3 h 的预处理。采用电化学技术、EDX 和 XPS 分析以及 HeLa 细胞系的细胞毒性试验,研究了处理前后铜的腐蚀行为和生物相容性。结果表明,PT 和 6-MP 溶液处理 3 和 1 h 的铜样比未经处理的铜样更具生物相容性。此外,该处理减少了放置后第一周内模拟子宫溶液中铜的爆发式释放效应。综上所述,PT 和 6-MP 处理是减少与铜基 IUD 相关的“爆发释放”副作用的有前途的策略,而不会改变其避孕作用。