Song Lipei, Elson Daniel S
Department of Surgery and Cancer and Hamlyn Centre for Robotic Surgery, Imperial College London, Exhibition Road, SW7 2AZ, London, UK.
Biomed Opt Express. 2013 Jan 1;4(1):89-104. doi: 10.1364/BOE.4.000089. Epub 2012 Dec 13.
Laser speckle contrast analysis (LASCA) is limited to being a qualitative method for the measurement of blood flow and tissue perfusion as it is sensitive to the measurement configuration. The signal intensity is one of the parameters that can affect the contrast values due to the quantization of the signals by the camera and analog-to-digital converter (ADC). In this paper we deduce the theoretical relationship between signal intensity and contrast values based on the probability density function (PDF) of the speckle pattern and simplify it to a rational function. A simple method to correct this contrast error is suggested. The experimental results demonstrate that this relationship can effectively compensate the bias in contrast values induced by the quantized signal intensity and correct for bias induced by signal intensity variations across the field of view.
激光散斑对比分析(LASCA)仅限于作为一种测量血流和组织灌注的定性方法,因为它对测量配置敏感。由于相机和模数转换器(ADC)对信号进行量化,信号强度是会影响对比值的参数之一。在本文中,我们基于散斑图案的概率密度函数(PDF)推导了信号强度与对比值之间的理论关系,并将其简化为有理函数。提出了一种校正这种对比误差的简单方法。实验结果表明,这种关系可以有效补偿量化信号强度引起的对比值偏差,并校正视场中信号强度变化引起的偏差。