Yoon Jung-Ro, Jeong Hyeon-Il, Oh Kwang-Jun, Yang Jae-Hyuk
Department of Orthopaedic Surgery, Veterans Health Service Medical Center, Seoul, Republic of Korea.
Korea Orthopedic Clinic, Seoul, Republic of Korea.
Knee. 2014 Jan;21(1):318-21. doi: 10.1016/j.knee.2012.12.014. Epub 2013 Jan 20.
We report a case of polyethylene insert breakage in a 45-year-old man after 3.5 years of cruciate retaining type mobile bearing total knee arthroplasty (TKA). Interstingly, both condyles of the polyethylene insert have fractured. The visual assessment done by stereoscopic microscope in the investigation report suggested that the fracture propagation was a result of cyclic loading and that the fracture was from the articular surface as a result of tibio-femoral and anteroposterior shear loading. The initial flexion-extension gap mismatch and/or specific Asian habits like kneeling or deep knee bending could have been the possible factors for over-stress for the insert causing this complication. After replacement of the broken insert and modification for daily activity preventing deep knee flexion, the patient obtained complete relief of previous symptoms. It remains unclear whether insert breakage was secondary to polyethylene insufficient design or to the polyethylene material fracture propagation.
我们报告一例45岁男性患者,在接受十字韧带保留型活动轴承全膝关节置换术(TKA)3.5年后出现聚乙烯衬垫破裂的病例。有趣的是,聚乙烯衬垫的两个髁都发生了骨折。调查报告中通过立体显微镜进行的视觉评估表明,骨折的扩展是循环载荷的结果,并且骨折是由于胫股和前后剪切载荷导致从关节表面开始的。最初的屈伸间隙不匹配和/或特定的亚洲习惯,如跪姿或深度屈膝,可能是导致衬垫过度应力从而引发这种并发症的可能因素。在更换破裂的衬垫并对日常活动进行调整以防止深度屈膝后,患者先前的症状完全缓解。目前尚不清楚衬垫破裂是继发于聚乙烯设计不足还是聚乙烯材料的骨折扩展。