[Detection of monoamine neurotransmitters and its metabolites by high performance liquid chromatograph after pre-column derivatization of dansyl chloride column].

作者信息

Huang Xiao, Chen Jia-wen, He Li-ping, Kang Xue-jun

机构信息

Department of Preventive Medicine and Medical Examination, Xiangnan University, Chenzhou 423000, China.

出版信息

Zhonghua Yu Fang Yi Xue Za Zhi. 2012 Dec;46(12):1117-20.

Abstract

OBJECTIVE

To develop a high performance liquid chromatography (HPLC) for detection of monoamine neurotransmitters and its metabolites after pre-column derivatization with dansyl chloride.

METHODS

The C(18) chromatograph column (150 mm×4.6 mm×5 µm) was selected for detection, and derived by dansyl chloride (10 mg/ml) under the condition of 50°C water bath by pH11 buffer solution. 20 µl acetic acid acetone solution (1.0 mol/L) was then mixed in for termination of the reaction. Then the solution was cooling to room temperature, 0.1 mol/L acetic acid zinc-acetonitrile-tetrahydrofuran solution was adopted for mobile phrase, with the volume ratio at 62:35:3. The flow rate was 1.0 ml/min between 0-10 min, 2.0 ml/min between 10-35 min. The ultraviolet detection wavelength was 286 nm. The above method separately detected monoamine neurotransmitters and its metabolites and evaluated the limit of detection, accurate degree and accuracy degree.

RESULTS

The linear relations between each component was good in the range of 1 - 20 µg/ml (r = 0.999). The lowest detection limit of norepinephrine, dopamine, 5-hydroxytryptamine and the metabolites 3-methoxy-4-benzoglycols, homovanillic acid and 5-heteroauxin were separately 0.60, 0.80, 0.41, 0.21, 0.19 and 0.1 µg/ml; while the average recovery rates were between 78.5% - 95.9%, and the relative standard deviation (RSD) was 6.62%, 7.64%, 2.98%, 3.60%, 5.09% and 3.09%, respectively. In the process of selection and optimization of the chromatographic conditions, we observed the importance of metal ions to discretion, and discussed the temperature, pH of the buffer solution and dosage of dansyl chloride in derivation. Under the above conditions, the reaction was perfect, and the baseline of the detected materials thoroughly separated.

CONCLUSION

The method to detect monoamine neurotransmitters and its metabolites by HPLC and pre-column derivatization with dansyl chloride was established; and this method could provide reference for the detection of polyamine by HPLC.