一种用于动态心电图监测中减少运动伪影的自适应滤波和独立成分分析的优化数字信号处理器实现方法。

An optimized DSP implementation of adaptive filtering and ICA for motion artifact reduction in ambulatory ECG monitoring.

作者信息

Berset Torfinn, Geng Di, Romero Iñaki

机构信息

Holst Cenhe/imec, High Tech Campus 31, 5656AE Eindhoven, The Netherlands.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:6496-9. doi: 10.1109/EMBC.2012.6347482.

Abstract

Noise from motion artifacts is currently one of the main challenges in the field of ambulatory ECG recording. To address this problem, we propose the use of two different approaches. First, an adaptive filter with electrode-skin impedance as a reference signal is described. Secondly, a multi-channel ECG algorithm based on Independent Component Analysis is introduced. Both algorithms have been designed and further optimized for real-time work embedded in a dedicated Digital Signal Processor. We show that both algorithms improve the performance of a beat detection algorithm when applied in high noise conditions. In addition, an efficient way of choosing this methods is suggested with the aim of reduce the overall total system power consumption.

摘要

运动伪影产生的噪声是目前动态心电图记录领域的主要挑战之一。为解决这一问题,我们提出使用两种不同的方法。首先,描述了一种以电极 - 皮肤阻抗作为参考信号的自适应滤波器。其次,引入了一种基于独立成分分析的多通道心电图算法。这两种算法均针对嵌入专用数字信号处理器的实时工作进行了设计和进一步优化。我们表明,在高噪声条件下应用时,这两种算法均能提高心跳检测算法的性能。此外,还提出了一种选择这些方法的有效方式,目的是降低整个系统的总功耗。