Morrison Alexander M, Liang Tao, Douberly Gary E
Department of Chemistry, University of Georgia, Athens, Georgia 30602, USA.
Rev Sci Instrum. 2013 Jan;84(1):013102. doi: 10.1063/1.4776179.
We report the automation of a continuous-wave, singly resonant, optical parametric oscillator (Lockheed-Martin Aculight ARGOS 2400-SF-15). This commercially available optical parametric oscillator (OPO) is capable of producing >1 W of continuously tunable idler output between 2.2 and 4.6 μm. An algorithm based on the feedback from a high accuracy wavemeter is implemented to synchronize three separate OPO tuning elements; the translation of a fan-out type periodically poled lithium niobate crystal, the rotation of an intracavity etalon, and the continuous tuning of the pump and idler wavelengths via piezoelectric strain of the tunable fiber pump laser. This allows for several hundred wavenumbers of efficient, automatic, continuous tuning of the idler wave. Continuous feedback from the wavemeter limits the absolute frequency accuracy to ±20 MHz. The broad, automatic tuning of the OPO is demonstrated via its implementation as a probe laser for the infrared action spectroscopy of methanol solvated in helium nanodroplets. LabVIEW virtual instruments for the automation of this OPO laser system are reported, along with detailed schematics of the associated hardware developed at the University of Georgia.
我们报告了一种连续波、单共振光学参量振荡器(洛克希德·马丁公司的Aculight ARGOS 2400-SF-15)的自动化情况。这种商用光学参量振荡器(OPO)能够在2.2至4.6μm之间产生大于1W的连续可调闲频输出。基于高精度波meter反馈的算法被用于同步三个独立的OPO调谐元件;扇出型周期极化铌酸锂晶体的平移、腔内标准具的旋转以及通过可调谐光纤泵浦激光器的压电应变对泵浦和闲频波长进行连续调谐。这使得闲频波能够进行数百波数的高效、自动、连续调谐。来自波meter的连续反馈将绝对频率精度限制在±20MHz。通过将OPO用作氦纳米液滴中甲醇红外作用光谱的探测激光器,展示了其广泛的自动调谐功能。报告了用于该OPO激光系统自动化的LabVIEW虚拟仪器,以及佐治亚大学开发的相关硬件的详细原理图。