J Perinat Med. 2013 Jul;41(4):421-7. doi: 10.1515/jpm-2012-0180.
To evaluate the intra-operator reliability of tissue Doppler imaging-based quantitative cervical elastography in at-term pregnancies.
Three series of two cycles of manual gentle compression-relaxation of the cervix were performed by one gynecologist in 11 consecutive at-term pregnancies through the vaginal probe. The compression movements should be sufficient to obtain the maximal compression of the anterior portion of the cervical tissue, until the posterior portion begins to be dislocated. Strain values were assessed on the entire thickness of the anterior cervical lip. The influence of strain type (Lagrangian vs. natural) and the extent of compression exerted (difference among the two cycles of compression-relaxation) were evaluated.
Strain measurement was influenced by the extent of compression, as well as by the strain type. The strain measured during the cycle with larger cervical compression, using the natural strain preset, showed a superior reliability [mean strain among patients: 0.68±0.18; mean of differences among three measures: 0.07±0.06; intraclass correlation coefficient (ICC) consistency: 0.90]. The strain measured using the Lagrangian strain preset showed overall a low reliability (ICC consistency: 0.04).
Quantitative cervical elastography performed in at-term pregnancies, under standardized conditions, has a high reliability.
评估基于组织多普勒成像的定量宫颈弹性成像在足月妊娠中的操作者内可靠性。
一名妇科医生通过阴道探头对 11 例连续足月妊娠进行了 3 组 2 个周期的手动轻柔宫颈压迫-放松。压迫动作应足以获得宫颈组织前部分的最大压迫,直到后部分开始移位。在整个前宫颈唇厚度上评估应变值。评估了应变类型(拉格朗日与自然)和施加的压缩程度(两个压缩-放松周期之间的差异)的影响。
应变测量受压缩程度以及应变类型的影响。使用自然应变预设,在具有较大宫颈压缩的周期中测量的应变显示出更高的可靠性[患者平均应变:0.68±0.18;三次测量之间的差异平均值:0.07±0.06;组内相关系数(ICC)一致性:0.90]。使用拉格朗日应变预设测量的应变总体上可靠性较低(ICC 一致性:0.04)。
在标准化条件下,在足月妊娠中进行的定量宫颈弹性成像具有很高的可靠性。