圆窗电极插入与传统耳蜗造口术比较后的人工耳蜗性能比较。

Comparison of cochlear implant performance after round window electrode insertion compared with traditional cochleostomy.

机构信息

New York Medical College, Valhalla, New York, USA.

出版信息

Otolaryngol Head Neck Surg. 2013 May;148(5):822-6. doi: 10.1177/0194599813479576. Epub 2013 Feb 27.

Abstract

OBJECTIVE

To compare speech perception after undergoing round window (RW) electrode insertion with those who underwent traditional cochleostomy (C) cochlear implant (CI) surgery.

STUDY DESIGN

Case series with chart review.

SETTING

Academic cochlear implant center.

SUBJECTS

Patients who met CI criteria with favorable intraoperative RW anatomy who underwent RW CI electrode insertion. Patients who underwent traditional C surgery were matched for age, duration of deafness, and preoperative hearing.

METHODS

Postoperative speech performance using the consonant-nucleus-consonant (CNC) test, Northwestern University Children's Perception of Speech (NUCHIPS), and Hearing in Noise Test (HINT) between 3 and 48 months after CI.

RESULTS

A total of 84 patients underwent RW insertion between 2004 and 2011. Of these, 55 had pre- and postoperative speech perception data for analysis. Average age of implantation ranged from 2.4 to 87.3 years (mean ± SD, 42.1 ± 24.2 years) for RW patients. Average age of implantation for matched traditional C patients ranged from 2.3 to 85.6 years (mean ± SD, 43.3 ± 25.0 years). No significant differences in postoperative CI speech perception scores were noted between the RW and C groups at 12 months post-CI (RW group, 55.28% ± 23.26% vs C group, 53.19% ± 24.14%; P = .70).

CONCLUSION

Our study found that patients with favorable RW anatomy who underwent RW CI electrode insertion demonstrated comparable speech perception compared with the traditional cochleostomy insertion group.

摘要

目的

比较行圆窗(RW)电极植入术和传统耳蜗造口术(C)人工耳蜗植入术(CI)患者的言语感知。

研究设计

病例系列和图表回顾。

设置

学术人工耳蜗植入中心。

受试者

符合 CI 标准且 RW 解剖结构良好的患者,行 RW CI 电极植入术。为年龄、耳聋持续时间和术前听力匹配行传统 C 手术的患者。

方法

术后 3 至 48 个月,采用辅音-核-辅音(CNC)测试、西北大学儿童言语感知(NUCHIPS)和噪声下听力测试(HINT)评估术后言语表现。

结果

2004 年至 2011 年共 84 例行 RW 插入术。其中 55 例行术前和术后言语感知数据分析。RW 患者的平均植入年龄为 2.4 至 87.3 岁(均值±SD,42.1±24.2 岁)。匹配的传统 C 患者的平均植入年龄为 2.3 至 85.6 岁(均值±SD,43.3±25.0 岁)。RW 和 C 组患者术后 12 个月 CI 言语感知评分无显著差异(RW 组,55.28%±23.26%比 C 组,53.19%±24.14%;P=.70)。

结论

我们的研究发现,RW 解剖结构良好且行 RW CI 电极植入术的患者与传统耳蜗造口术插入组患者的言语感知能力相当。